🌿 Ammonia (NH₃) in waste gases — manual absorption method

Environment PN-EN ISO 21877

In short

A representative gas sample is taken with a heated probe through a particle filter into two gas-washing bottles with 0. The test is performed according to PN-EN ISO 21877:2020-03 (ISO 21877:2019); Range of validation: about 8–65 mg/m³ NH₃ (standard conditions); lower concentrations depending on the sampling and the limit of detection. The procedure comprises 6 steps; it is used for: Measurement of ammonia slip downstream of SCR/SNCR plants in power stations, cement works and waste incineration plants, Emissions of ammonia from chemical works (the manufacture of fertilisers) and agricultural installations, Calibration and validation of fixed automated measuring systems for ammonia (ISO 17179, EN 14181).

At a glance

  • Standard: PN-EN ISO 21877:2020-03 (ISO 21877:2019)
  • Category: Environment
  • Procedure steps: 6
  • Range of validation: about 8–65 mg/m³ NH₃ (standard conditions); lower concentrations depending on the sampling and the limit of detection
  • Absorption solution: H₂SO₄ 0.05 mol/l of analytical grade (the concentration may be increased); demineralised water
  • Temperature of the line: at least 15 K above the dew point of the gas, as far as the gas-washing bottles

Overview

The emissions of ammonia come to a large extent from agriculture, from chemical processes (e.g. the manufacture of fertilisers), from power stations, cement works and waste incineration plants with SCR and non-SCR reactors (ammonia slip). ISO 21877:2019 specifies a manual measurement method — sampling and various analytical methods — for determining the mass concentration of ammonia in the waste gases of industrial plants, e.g. combustion plants or agricultural installations. The method determines all the compounds volatile at the temperature of sampling which during the absorption dissociate with the formation of ammonium ions — that is the volatile ammonia content of the gas. It is an independent measurement method (called in EN 14181 the standard reference method), validated in field tests in the range of about 8–65 mg/m³ under standard conditions; the lower limit was set by the operating conditions of the test plant, and the method may be applied at lower concentrations depending on the time and the volume of sampling and on the limit of detection of the analysis. It serves for periodic monitoring and for the calibration and the validation of fixed automated measuring systems for ammonia (in conjunction with ISO 17179). In Poland the standard is in force as PN-EN ISO 21877:2020-03.

Method principle

A representative sample is taken from the gas stream over a specified time at a specified flow; where there are droplets in the gas, isokinetic sampling is necessary. The probe is heated to a temperature ensuring the evaporation of the droplets and the absence of condensation of water vapour; the particles are retained by a specified filter — in-stack filtration (preferred with non-isokinetic sampling, because it does not require separate heating) or out-stack filtration, in which case the filter is heated to a set temperature. The ammonia passing through the filter is collected in an absorption system acidified with H₂SO₄ (a 0.05 M solution; the concentration may be increased at high concentrations of NH₃), and the mass of NH₄⁺ is determined after sampling by one of the analytical methods of clause 9 (spectrometry in accordance with ISO 7150-1, flow analysis CFA/FIA in accordance with ISO 11732, ion chromatography in accordance with ISO 14911). The volume of the gas is measured with a gas meter; the mass concentration is the quotient of the mass of ammonia in the solution (mg) and the volume of the gas (m³), with the conversion factor NH₄⁺→NH₃ fN = 0.944. The sampling system: a heated probe with an inlet of titanium, quartz, borosilicate glass or PTFE; a flat quartz filter; two gas-washing bottles with a sintered disc (porosity D1 or finer) at low flows, or impingers at high flows; a pump, a flow controller, a gas meter with a thermometer and a manometer, optionally a drying tower; all the components resistant to corrosion; the line to the gas-washing bottles heated to at least 15 K above the dew point of the gas, with the residence time minimised. In the presence of semi-volatile ammonium salts the temperature of sampling affects the gas/solid equilibrium; the analysis of the dust (total ammonia) is not covered by the standard.

Applications

Key parameters

ParameterValue
Range of validationabout 8–65 mg/m³ NH₃ (standard conditions); lower concentrations depending on the sampling and the limit of detection
Absorption solutionH₂SO₄ 0.05 mol/l of analytical grade (the concentration may be increased); demineralised water
Temperature of the lineat least 15 K above the dew point of the gas, as far as the gas-washing bottles
Gas-washing bottlestwo: gas-washing bottles with a sintered disc (bubblers) of D1 or finer (low flow) or impingers (high flow)
Analysis of NH₄⁺spectrometry (ISO 7150-1), CFA/FIA (ISO 11732) or ion chromatography (ISO 14911); fN = 0.944
Uncertaintyrelative expanded ≤ 20 % of the reference value (e.g. the daily ELV), for dry gas before the correction for O₂
Reference conditions273 K, 101.3 kPa; the result for dry gas and reference O₂

Standard

Standard number
PN-EN ISO 21877:2020-03 (ISO 21877:2019)
Title (PL)
Emisja ze źródeł stacjonarnych — Oznaczanie stężenia masowego amoniaku — Metoda manualna
Title (EN)
Stationary source emissions — Determination of the mass concentration of ammonia — Manual method

Step-by-step procedure

  1. Planning

    Establish the measurement site and the measurement points (EN 15259), the presence of droplets (→ isokinetic sampling), the expected concentration of NH₃ and the sampling time such that the mass of NH₄⁺ exceeds the limit of quantification of the analysis.

  2. Assembly of the system

    Assemble the heated probe, the filter (in-stack or heated), two gas-washing bottles with 0.05 M H₂SO₄, a safety bottle, a drying tower (with a dry gas meter), the pump, the flow meter and the gas meter; heat the line to the gas-washing bottles to at least 15 K above the dew point; check the leak-tightness.

  3. Sampling

    Sample over a specified time at a specified flow (with isokineticity — with a side line of a constant ratio of the flows, or without one); record the volume, the temperature and the pressure at the gas meter. Prepare a field blank.

  4. Recovery of the samples

    Pour the solutions from both gas-washing bottles separately into glass, PP or PE bottles, rinse the components downstream of the filter with demineralised water and add the rinsings; assess the effectiveness of the absorption from the share of the second gas-washing bottle.

  5. Analysis

    Determine NH₄⁺ by the chosen method (spectrometry, CFA/FIA or ion chromatography) against a calibration curve, taking the dilution and the blanks into account.

  6. Calculations

    Calculate the mass of NH₃ (fN = 0.944), the volume of gas under standard conditions, the mass concentration in mg/m³ for dry gas and after the correction for reference O₂; draw up an uncertainty budget (≤ 20 % of the reference value).

Required equipment and apparatus

EquipmentExampleIndicative price
Heated sampling probe with an inlet of titanium, quartz, borosilicate glass or PTFEwith a nozzle for isokinetic sampling—
Flat quartz filterin-stack (without heating) or out-stack in a heated housing—
Two gas-washing bottles with a sintered disc (D1 or finer) or impingerswith a 0.05 M H₂SO₄ solution; optionally a safety bottle—
Suction unit: pump, flow controller, gas meter with a thermometer and a manometer, drying tower (dry gas meter)flow meter downstream of the filter or upstream of the gas meter—
Equipment for isokineticitymeasurement of the pressure, the temperature and the composition of the gas; a main line and a side line with their own flow measuring systems (orifice plate)—
Spectrometer, flow analyser or ion chromatographin accordance with the analytical method chosen; glass, PP or PE bottles, pipettes, flasks of 50–1000 ml—

Reagents, media and consumables

ReagentCASDetails
Sulfuric acid 0.05 mol/l, analytical grade—absorption solution for the ammonia
Demineralised water—preparation of the solutions, rinsing
Standard solutions of NH₄⁺ and the reagents of the analytical method (in accordance with ISO 7150-1, ISO 11732 or ISO 14911)—calibration curve and determination of the ammonium ion

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN ISO 21877:2020-03 (ISO 21877:2019) — “Stationary source emissions — Determination of the mass concentration of ammonia — Manual method”.

How does this method work?

A representative sample is taken from the gas stream over a specified time at a specified flow; where there are droplets in the gas, isokinetic sampling is necessary. The probe is heated to a temperature ensuring the evaporation of the droplets and the absence of condensation of water vapour; the particles are retained by a specified filter — in-stack filtration (preferred with non-isokinetic sampling, because it does not require separate heating) or out-stack filtration, in which case the filter is heated to a set temperature.

What is the measuring range and accuracy?

Range of validation: about 8–65 mg/m³ NH₃ (standard conditions); lower concentrations depending on the sampling and the limit of detection; Absorption solution: H₂SO₄ 0.05 mol/l of analytical grade (the concentration may be increased); demineralised water; Temperature of the line: at least 15 K above the dew point of the gas, as far as the gas-washing bottles; Gas-washing bottles: two: gas-washing bottles with a sintered disc (bubblers) of D1 or finer (low flow) or impingers (high flow).

How long does the test take?

The procedure comprises 6 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Heated sampling probe with an inlet of titanium, quartz, borosilicate glass or PTFE, Flat quartz filter, Two gas-washing bottles with a sintered disc (D1 or finer) or impingers, Suction unit: pump, flow controller, gas meter with a thermometer and a manometer, drying tower (dry gas meter), Equipment for isokineticity, Spectrometer, flow analyser or ion chromatograph.

Where is this test used?

Measurement of ammonia slip downstream of SCR/SNCR plants in power stations, cement works and waste incineration plants; Emissions of ammonia from chemical works (the manufacture of fertilisers) and agricultural installations; Calibration and validation of fixed automated measuring systems for ammonia (ISO 17179, EN 14181); Periodic measurements for integrated permits and emission reporting; Parallel sampling with other components in the side lines of an isokinetic system.

What safety precautions apply?

Sulfuric acid solution — protection of the eyes and the hands; the solutions after sampling contain ammonium salts and may contain acid components of the flue gas; The heated probe and the filter are liable to cause burns; work on a stack platform with protection against falling; Ammonia in the gases and at leaking ports irritates the respiratory tract — protection of the respiratory tract when opening the ports.

Which laboratory can perform this test?