⚗️ AOX — adsorbable organically bound halogens in water and wastewater: adsorption on activated carbon, displacement of inorganic halides, combustion at 950 °C and argentometric titration

Physicochemistry PN-EN ISO 9562

In short

A summary indicator of the pollution of water and wastewater by organochlorine, organobromine and organoiodine compounds (fluorine is not included). The test is performed according to PN-EN ISO 9562:2007; Range: direct determination in water from usually 10 µg/l (ISO 9562:2004, clause 1 — the wording of the scope is abbreviated here and does not decide expressly whether 10 µg/l is the lower limit or a typical value; we give it as it stands). The procedure comprises 12 steps (about 45 min in total); it is used for: Control of industrial wastewater discharged directly and indirectly — demonstrating compliance with the branch AOX limit value, Assessment of the efficiency of a treatment plant: comparison of AOX in raw and treated wastewater, Monitoring of surface water and groundwater for organic halogen compounds.

At a glance

  • Standard: PN-EN ISO 9562:2007
  • Category: Physicochemistry
  • Procedure steps: 12
  • Total time of stages: 45 min
  • Quantity determined: the sum of organically bound chlorine, bromine and iodine (without fluorine), adsorbable on activated carbon, expressed as chlorides — an analytical convention, not the content of any particular compound
  • Range: direct determination in water from usually 10 µg/l (ISO 9562:2004, clause 1 — the wording of the scope is abbreviated here and does not decide expressly whether 10 µg/l is the lower limit or a typical value; we give it as it stands)
  • Chloride limitation: inorganic chlorides < 1 g/l; richer samples are diluted before analysis. For filtered samples of high salinity — the modified SPE-AOX method (Annex A, informative), whose results may differ significantly from those of the required method

Overview

AOX is not the name of a substance but an analytical convention — the standard says so expressly in its introduction. The result represents the sum of organically bound chlorine, bromine and iodine (but not fluorine) which under defined conditions can be adsorbed on activated carbon, and if the sample is not filtered it also covers the halogens bound to the suspended matter. The scope of the standard is the direct determination in water of usually 10 µg/l of organically bound chlorine, bromine and iodine, expressed as chlorides, adsorbable on activated carbon. The method applies to samples with an inorganic chloride concentration below 1 g/l; samples with higher concentrations are diluted before analysis. It also applies to samples with suspended matter on which halogens are adsorbed (e.g. insoluble halides) — filtering the sample before analysis makes it possible to determine dissolved AOX and AOX in the suspended matter separately. For filtered samples with a high content of inorganic chlorides the standard provides a modified method, with solid-phase extraction (SPE-AOX, Annex A, informative), with an explicit proviso: results obtained by the modified method may differ significantly from the results of the required method.

What it is for. The parameter serves above all for monitoring water quality, but it also plays a part in assessing solid waste (e.g. in aqueous eluates of waste) and sewage sludge. The recipients are municipal wastewater treatment plants, environmental authorities and industrial establishments discharging wastewater directly and indirectly. An industrial plant has to show that its wastewater does not exceed the AOX limit value proper to its branch; a treatment plant checks not only compliance with limit values but also the efficiency of its own process — by comparing AOX before and after treatment. That is the practical value of this indicator: a single number says how much halogen-bound organic matter there is in the wastewater, without deciding which compounds these are.

Three routes of adsorption and the choice between them. The standard describes three procedures of adsorption on activated carbon: shaking, stirring and column adsorption; the choice is imposed by laboratory or administrative rules, and in many countries the column method is obligatory for wastewater. The stirring method (also called carbodisc, with a carbon fibre fleece) requires very experienced staff and a great deal of manual work, which is why it is rare in routine laboratories, and it is poorly suited to samples with solid particles — typical of wastewater; the standard itself warns that with samples containing suspended matter, stirring may not cover the particles sufficiently and then recommends shaking or a column. The shaking (batch) method is obligatory for sewage sludge and other solid samples and is sometimes used for waters, but it has limited automation, a greater risk of contamination from the laboratory atmosphere and offers no simple check on the completeness of adsorption; it is not recommended for waters in which chlorides cannot be brought down by dilution to a value ≤ 500 mg/l Cl⁻. The column method covers samples with and without particles, permits a simple check on the completeness of adsorption (a second column in series), is more resistant to high chloride concentrations thanks to more efficient washing of the carbon with the nitrate solution, and lends itself to automation. The standard itself adds that at high chloride concentrations (about 1 g/l) shaking may give greater interferences than the column procedure.

What to expect from interferences. High AOX values may arise from the presence of active chlorine and of certain inorganic bromine and iodine compounds, which bind to the carbon irreversibly — the reactions of these oxidants with the organic compounds of the sample and with the carbon are prevented by adding sodium sulfite immediately after the sample is taken. Organic bromine and iodine compounds may decompose during combustion into elemental bromine or iodine and pass into higher oxidation states, so that they are sometimes determined incompletely (negative error). Samples containing living cells (microorganisms, algae) may give overestimated (falsely elevated) results because of their chloride content — such samples are not analysed earlier than 8 h after acidification. Alcohols, aromatic compounds and carboxylic acids may lower the result (e.g. at DOC > 100 mg/l). The recovery of some polar and hydrophilic compounds, such as monochloroacetic acid, is incomplete.

Status of the standard. The current Polish edition is PN-EN ISO 9562:2007 — the Polish-language version, published on 28 June 2007, 26 pages, price group M, managed by KT 122 for Water Quality — Chemical Tests — Organic Substances, ICS 13.060.50. It implements EN ISO 9562:2004 [IDT] and ISO 9562:2004 [IDT] and supersedes PN-EN ISO 9562:2005 — the English-language version. The PKN catalogue card for this edition has neither a "Date of withdrawal" field, nor a "Superseded by" field, nor a corrigendum (Ap) — the edition is in force. In the PKN list for number 9562 there are only two items: the withdrawn English version of 2005 and the binding Polish version of 2007, so the trap of "the corrigendum hangs only on one language version" does not occur here — only one version has a current card. On the international side ISO 9562:2004 is the third edition which — as its own foreword states — cancelled and replaced the technically revised second edition ISO 9562:1998; the first edition ISO 9562:1989, which we use below in a few expressly marked places, is thereby doubly out of date. On the European side EN ISO 9562:2004 superseded EN 1485:1996.

What this description does not contain. We have not seen the full text of the standard — PKN and ISO supply it for a fee. The scope, interferences, principle, list of reagents and apparatus (clauses 1–7; clause 8 "Sampling and pretreatment of samples" is also outside the preview) are given from the preview of ISO 9562:2004, and the Polish title and the publication data from the PKN catalogue card. Clause 9 (procedure), 10 (calculation), 11 (precision) and 12 (test report) and the annexes are not in the preview. That is why we give the working parameters of the column method from the Analytik Jena application notes for the multi X 2500 analyser, declared as compliant with DIN EN ISO 9562 (this is a manufacturer's source, not the text of the standard), and a few figures — expressly marked — from the preview of the withdrawn edition ISO 9562:1989, which in those places agrees with the manufacturer's notes. We give no data at all on the precision of the method from the current edition, because we have not seen them.

Method principle

The water sample is acidified with nitric acid. The organic compounds contained in the sample are adsorbed on activated carbon — by shaking, by stirring or by column adsorption. The adsorbed inorganic halides are displaced by washing the carbon with a sodium nitrate solution acidified with nitric acid. The loaded carbon is burned in a stream of oxygen, the hydrogen halides formed are absorbed in an absorption solution, and the halide ions are determined by argentometric titration, for example microcoulometrically. The result is expressed as the mass concentration of chlorides. The distinction between dissolved AOX and AOX bound to suspended matter does not follow from the chemistry of the determination but from whether the sample was filtered before analysis.

Applications

Key parameters

ParameterValue
Quantity determinedthe sum of organically bound chlorine, bromine and iodine (without fluorine), adsorbable on activated carbon, expressed as chlorides — an analytical convention, not the content of any particular compound
Rangedirect determination in water from usually 10 µg/l (ISO 9562:2004, clause 1 — the wording of the scope is abbreviated here and does not decide expressly whether 10 µg/l is the lower limit or a typical value; we give it as it stands)
Chloride limitationinorganic chlorides < 1 g/l; richer samples are diluted before analysis. For filtered samples of high salinity — the modified SPE-AOX method (Annex A, informative), whose results may differ significantly from those of the required method
Dissolved versus suspended AOXthey are separated solely by filtering the sample before analysis; an unfiltered sample gives a combined result
Combustion temperaturea furnace capable of at least 950 °C, a quartz tube of internal diameter 2–4 cm and length about 30 cm (ISO 9562:2004, 7.2.1)
Detectionan argentometric device, usually a microcoulometer, able to determine 1 µg of chlorides (in absolute terms) with a coefficient of variation of 10 % (repeatability)
Quality of the activated carboniodine number > 1050 according to the method from the bibliography of the standard; blank value of the washed carbon < 15 µg chloride equivalent per gram of carbon
Check on the completeness of adsorption (column)the AOX from the second column should not exceed 10 % of the value from the first; otherwise the sample is diluted — a figure from the withdrawn edition ISO 9562:1989, repeated by the manufacturer’s notes
Agreement of dilutionsaccording to the Analytik Jena note to DIN EN ISO 9562 the results of two different dilutions of the same sample should not differ by more than 10 %

Standard

Standard number
PN-EN ISO 9562:2007
Title (PL)
Jakość wody — Oznaczanie adsorbowalnych, organicznie związanych chlorowców (AOX)
Title (EN)
Water quality — Determination of adsorbable organically bound halogens (AOX)

Step-by-step procedure

  1. Taking and preserving the sample

    Take the sample into a flat-bottomed 1000 ml bottle, preferably of brown glass. Immediately after sampling add the sodium sulfite solution — without it active chlorine and oxidising bromine and iodine compounds will raise the result, binding irreversibly to the carbon. Samples containing living cells (microorganisms, algae) are analysed not earlier than 8 h after acidification.

    🌡 Temperature: 0–4 °C for storage of the sample

  2. Deciding: total or dissolved AOX

    If the result is to cover the halogens bound to the suspended matter — do not filter the sample. If dissolved and suspended AOX are to be separated — filter the sample before analysis and state this in the test report. This decision changes the meaning of the result, not merely its value.

  3. Checking the chloride concentration and diluting

    The method applies to samples with an inorganic chloride concentration below 1 g/l; richer samples are to be diluted before analysis. For saline and filtered waters the standard provides the modified SPE-AOX method (Annex A, informative), with the proviso that its results may differ significantly.

  4. Choosing the adsorption procedure

    Choose shaking, stirring or the column — laboratory or administrative rules decide. For wastewater the column method is obligatory in many countries; for sewage sludge and other solid samples — the shaking method. For samples with suspended matter the standard advises against stirring (the particles are sometimes insufficiently covered), and at chlorides of the order of 1 g/l — against shaking (greater interferences than on a column).

  5. Acidification and addition of nitrate (column method)

    According to the manufacturer’s note compliant with DIN EN ISO 9562: bring the sample to pH 2 with concentrated nitric acid and then add 5 ml of the nitrate stock solution to 100 ml of the prepared sample. Samples with a high particle content are to be homogenised before adsorption with a high-speed disperser.

  6. Adsorption on the columns

    Pass the sample through two adsorption tubes set in series, each filled with about 50 mg of activated carbon, at a flow rate of 3 ml/min. With samples containing particles a filtration column may be placed before the carbon columns — the particles are retained and the carbon columns do not block.

    ⏱ Time: about 45 min for six samples on a parallel unit according to the manufacturer’s note

  7. Washing the carbon with the nitrate solution

    Wash the columns with the nitrate washing solution at the same rate of 3 ml/min — according to the manufacturer’s note 25 ml per determination. This step displaces the inorganic halides from the carbon; without it the inorganic chlorides would be counted as AOX.

  8. Combustion

    Burn the loaded carbon in a stream of pure oxygen at a temperature of at least 950 °C. The columns are introduced into a vertical quartz tube (in the arrangement described in the manufacturer’s note this is done by an automatic feeder). The organohalogen compounds are converted into hydrogen halides.

    🌡 Temperature: ≥ 950 °C

  9. Drying the gas and detection

    Dry the combustion gas with concentrated sulfuric acid in an absorber and then introduce it into the microcoulometric measuring cell. The chlorides are determined by argentometric titration. Microcoulometry is an absolute method, so according to the manufacturer’s note the analyser does not require calibration — it does require confirmation of correct operation with an AOX standard solution (4-chlorophenol in water).

  10. Checking the completeness of adsorption and the agreement of dilutions

    Read the result from the first and from the second column separately. According to the withdrawn edition ISO 9562:1989 the AOX value from the second column should not exceed 10 % of the value from the first — otherwise the sample should be diluted. According to the manufacturer’s note to DIN EN ISO 9562 the results of two different dilutions of the same sample should not differ by more than 10 %; that is the simplest test of whether the matrix is overloading the adsorption.

  11. Blanks and checking the procedure

    Determine the total blank — it covers the water, the reagents and the gases. According to the edition ISO 9562:1989 the total blank should not exceed 3 µg (or 30 µg/l respectively), and if it does, the adsorption, the combustion and the microcoulometric determination are checked separately. The same edition requires, when checking the whole procedure, five standard solutions and a blank, and regards the result as acceptable when the correlation coefficient is greater than 0.999; the manufacturer’s note adds the criterion of a slope of 0.95–1.05. Note: these figures come from outside the current edition — before use check the wording of clauses 9 and 10 in the copy of PN-EN ISO 9562:2007 you hold.

  12. Calculation and test report

    Express the result as the mass concentration of chlorides (µg/l). In the test report state whether the sample was filtered (dissolved or total AOX), which adsorption procedure was used, what dilution was applied and what the blank was. Without this information the result of an analytical convention is not comparable with any other.

Required equipment and apparatus

EquipmentExampleIndicative price
Combustion apparatus with furnace and quartz tubea furnace heated to at least 950 °C, a quartz tube of internal diameter 2–4 cm and length about 30 cm; both vertical and horizontal tubes are met with—
Argentometric measuring device (microcoulometer)able to determine 1 µg of chlorides in absolute terms with a coefficient of variation of 10 %; titration with a dilute AgNO₃ solution is also admissible—
Absorber with sulfuric acid for drying the gasthe acid must not flow back into the furnace; it is replaced when its volume increases by more than 20 %—
Set for column adsorptiona pump (e.g. a piston pump) with PTFE tubing and adsorption tubes of internal diameter about 3 mm and length 40–50 mm, set vertically one behind the other, each with about 50 mg of activated carbon 50–150 µm, held in place with halogen-free ceramic wool—
Set for adsorption by shakingfiltration apparatus (a funnel of 0.15 l capacity and filter diameter 25 mm), a polycarbonate membrane filter 0.45 µm, 250 ml conical flasks according to ISO 1773, a mechanical shaker with a carrier plate; activated carbon with a grain size of about 10–50 µm—
Set for adsorption by stirringa 100 ml conical flask, a carbon fibre fleece (e.g. discs 13 mm × 3 mm, 100 % carbon, specific surface 1600 m²/g, mass 25 mg — two fleeces per determination, 50 mg in total), a magnetic stirrer with a funnel for washing—
Sampling bottlesflat-bottomed, 1000 ml, preferably of brown glass; for checking the microtitration a syringe for 10–100 µl of hydrochloric acid—

Reagents, media and consumables

ReagentCASDetails
Grade 1 water—according to ISO 3696:1987; the AOX content of the water, the reagents and the gases must be negligible in relation to the lowest AOX content determined, which is checked by a total blank
Activated carbon—for shaking a grain size of about 10–50 µm, for the column about 50–150 µm, for stirring — a commercial carbon fleece. Iodine number > 1050; blank value of the washed carbon < 15 µg chloride equivalent per gram
Concentrated nitric acid7697-37-2ρ(HNO₃) = 1.4 g/ml, w(HNO₃) = 65 %; for acidifying the sample and for preparing the nitrate solutions. Separately, dilute nitric acid c(HNO₃) = 0.02 mol/l
Sodium nitrate — stock solution c(NaNO₃) = 0.2 mol/l7631-99-4dissolve 17 g of NaNO₃ in water in a 1000 ml flask, add 25 ml of concentrated HNO₃ and make up to the mark; stable for three months in a brown bottle
Nitrate washing solution c(NaNO₃) = 0.01 mol/l, pH ≈ 1.7—make up 50 ml of the stock solution with water to 1000 ml; stable for a month in a brown bottle. It serves to displace the inorganic halides from the carbon
Sodium sulfite — solution c(Na₂SO₃) = 1 mol/l7757-83-7dissolve 126 g of Na₂SO₃ and make up with water to 1000 ml; stable for a month at 2–8 °C. It is added immediately after the sample is taken so that active chlorine and oxidising bromine and iodine compounds do not react with the organic compounds or with the carbon
Hydrochloric acid c(HCl) = 0.010 mol/l7647-01-0the concentration must be exactly known — the acid serves to check the microtitration
Sulfuric acid7664-93-9ρ(H₂SO₄) = 1.84 g/ml; it fills the absorber drying the gas before the measuring cell
4-Chlorophenol — stock solution ρCl(AOX) = 200 mg/l106-48-972.5 mg of 4-chlorophenol (C₆H₅ClO) in a 100 ml flask; the working solution of 1 mg/l is made from 5 ml of the stock made up to 1000 ml. The stock is kept for a month, the working solution for a week, at 2–8 °C in glass. For safety reasons the standard recommends buying ready-made solutions
2-Chlorobenzoic acid — stock solution ρCl(AOX) = 250 mg/l118-91-2110.4 mg of the acid (ClC₆H₄COOH) in a 100 ml flask; it dissolves very slowly, which is why the standard recommends preparing the solution a day in advance. Working solution 1 mg/l: 4 ml of the stock made up to 1000 ml
Standard solutions for checking the whole procedure—from the working solutions of 4-chlorophenol or 2-chlorobenzoic acid: e.g. 1, 5, 10, 20 and 25 ml into 100 ml flasks, giving AOX of 10, 50, 100, 200 and 250 µg/l. The concentrations are chosen so as to cover the whole working range; they are prepared on the day of use
Oxygen for combustion—oxygen or a mixture of oxygen with an inert gas; the manufacturer’s note gives oxygen of 99.995 % purity (4.5)
Potassium iodide7681-11-0a reagent named in the list of the standard (clause 6.15 or 6.16 — the preview names both numbers but does not decide which refers to which reagent)
Starch solution 1 %—an indicator solution named in the list of the standard next to potassium iodide (clause 6.15 or 6.16). Without a CAS number: starch is a polymer of variable composition and has no single number for the substance

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN ISO 9562:2007 — “Water quality — Determination of adsorbable organically bound halogens (AOX)”.

How does this method work?

The water sample is acidified with nitric acid. The organic compounds contained in the sample are adsorbed on activated carbon — by shaking, by stirring or by column adsorption.

What is the measuring range and accuracy?

Quantity determined: the sum of organically bound chlorine, bromine and iodine (without fluorine), adsorbable on activated carbon, expressed as chlorides — an analytical convention, not the content of any particular compound; Range: direct determination in water from usually 10 µg/l (ISO 9562:2004, clause 1 — the wording of the scope is abbreviated here and does not decide expressly whether 10 µg/l is the lower limit or a typical value; we give it as it stands); Chloride limitation: inorganic chlorides < 1 g/l; richer samples are diluted before analysis. For filtered samples of high salinity — the modified SPE-AOX method (Annex A, informative), whose results may differ significantly from those of the required method; Dissolved versus suspended AOX: they are separated solely by filtering the sample before analysis; an unfiltered sample gives a combined result.

How long does the test take?

The times given for the individual stages add up to approximately 45 min. The procedure comprises 12 steps.

What equipment is required?

Combustion apparatus with furnace and quartz tube, Argentometric measuring device (microcoulometer), Absorber with sulfuric acid for drying the gas, Set for column adsorption, Set for adsorption by shaking, Set for adsorption by stirring, Sampling bottles.

Where is this test used?

Control of industrial wastewater discharged directly and indirectly — demonstrating compliance with the branch AOX limit value; Assessment of the efficiency of a treatment plant: comparison of AOX in raw and treated wastewater; Monitoring of surface water and groundwater for organic halogen compounds; Control of waters after chlorine disinfection — chlorinated by-products as a summary indicator (with attention to the addition of sodium sulfite binding active chlorine); Testing of aqueous eluates of solid waste and sewage sludge (for solids the shaking method is obligatory); Pulp-and-paper and chemical wastewater — the edition ISO 9562:1989 names wastewater from pulp mills as requiring longer adsorption because of high-molecular constituents, and the manufacturer’s note examines raw wastewater from a chemical plant.

What safety precautions apply?

Concentrated nitric acid (65 %) and concentrated sulfuric acid are strongly corrosive and oxidising. Acidify samples and fill the absorber in a fume cupboard, wearing acid-resistant gloves, goggles and a laboratory coat.; For safety reasons the standard itself recommends buying ready-made 4-chlorophenol standard solutions instead of preparing them from the solid substance — the simplest way of avoiding contact with the reagent during weighing.; The combustion furnace works at a temperature of at least 950 °C with a flow of pure oxygen — the combination of high temperature and oxygen requires keeping combustible substances away and checking the tightness of the oxygen installation.; The sulfuric acid from the absorber must not flow back into the heated quartz tube when the temperature or the pressure falls — this leads to violent evaporation and cracking of the tube.; Laboratory air is sometimes appreciably contaminated with organic halogen compounds (reagents, solvents, correction fluid) — this is not only a problem of the result but also of exposure; the carbon on the filter must not be sucked dry, which reduces both the error and the contact with the atmosphere.; Industrial wastewater tested for AOX is by assumption sometimes loaded with organohalogen compounds of carcinogenic and mutagenic action — treat the sample as hazardous material and send the waste from adsorption and combustion for chemical disposal.

Which laboratory can perform this test?