🎨 Synthetic Food Dyes (HPLC)

Food Chemistry PN-EN 12878

In short

Identification and quantitative determination of synthetic food dyes (tartrazine, sunset yellow, azorubine, allura red etc. The test is performed according to PN-EN 12878:2014 (oraz metody wewnętrzne oparte na HPLC-DAD); Determination range: 0.5–500 mg/kg (depending on dye and matrix). The procedure comprises 10 steps (about 1 h 45 min in total); it is used for: Official food control by Sanitary Inspection (SANEPID), Quality control of carbonated drinks, juices, syrups, Testing of sweets, candies, jellies for azo dyes.

At a glance

  • Standard: PN-EN 12878:2014 (oraz metody wewnętrzne oparte na HPLC-DAD)
  • Category: Food Chemistry
  • Procedure steps: 10
  • Total time of stages: 1 h 45 min
  • Determination range: 0.5–500 mg/kg (depending on dye and matrix)
  • Limit of detection (LOD): 0.1–0.5 mg/kg
  • Limit of quantification (LOQ): 0.5–1.0 mg/kg

Overview

Synthetic food dyes (E100–E199) are among the most frequently controlled food additives due to potential health risks — allergic reactions, hyperactivity in children (so-called "Southampton six") and possible carcinogenic effects of some azo dyes. Regulation (EC) 1333/2008 specifies permitted dyes and their maximum doses in individual food categories.

The HPLC method with UV-Vis (DAD — diode array detector) detection is the reference analytical technique for simultaneous identification and quantitative determination of many synthetic dyes in one analysis. Reversed-phase chromatography (RP-HPLC) on a C18 column with gradient elution allows separation and quantification of all permitted synthetic dyes in approximately 20–30 minutes.

Most commonly determined dyes are: tartrazine (E102), quinoline yellow (E104), sunset yellow (E110), azorubine (E122), ponceau 4R (E124), allura red (E129), brilliant blue FCF (E133), indigotine (E132) and green S (E142). The method requires preliminary extraction of dyes from the food matrix, purification on SPE cartridges and chromatographic analysis.

Accredited laboratories performing these tests participate in proficiency testing programs (PT) organized by FAPAS and LGC, ensuring reliability of results at the international level.

Method principle

Synthetic dyes are extracted from a food sample with a mixture of water/methanol/ammonia, purified on SPE cartridges (polyamide or C18) and separated by liquid chromatography on a C18 column with gradient elution (mobile phase: ammonium acetate buffer pH 7 / acetonitrile). Detection using a DAD (diode array detector) at multiple wavelengths (427 nm, 482 nm, 520 nm, 630 nm) allows dye identification based on absorption spectrum and retention time. Quantitative determination — from calibration curve of standards.

Applications

Key parameters

ParameterValue
Determination range0.5–500 mg/kg (depending on dye and matrix)
Limit of detection (LOD)0.1–0.5 mg/kg
Limit of quantification (LOQ)0.5–1.0 mg/kg
ColumnC18 (250 × 4.6 mm, 5 µm) or UHPLC C18 (100 × 2.1 mm, 1.8 µm)
Detection wavelengths427, 482, 520, 630 nm (DAD: full spectrum 200–700 nm)
Analysis time20–30 min (HPLC), 8–12 min (UHPLC)

Standard

Standard number
PN-EN 12878:2014 (oraz metody wewnętrzne oparte na HPLC-DAD)
Title (PL)
Oznaczanie syntetycznych barwników w żywności metodą HPLC
Title (EN)
Determination of synthetic food colorants by HPLC

Step-by-step procedure

  1. Sample preparation

    Homogenize food sample (blender). Weigh 2–5 g into volumetric flask.

    ⏱ Time: 10 min

  2. Dye extraction

    Add 20 mL of methanol/water/ammonia mixture (70:25:5). Place in ultrasonic bath for 15 min. Filter.

    ⏱ Time: 20 min

  3. SPE cleanup

    Condition SPE cartridge (5 mL methanol, 5 mL water). Load extract. Rinse 5 mL water. Elute dyes with 5 mL methanol with ammonia.

  4. Concentration and dissolution

    Evaporate eluent to dryness under nitrogen stream (40°C). Dissolve residue in 1 mL mobile phase A. Filter through 0.22 µm into vial.

    ⏱ Time: 15 min • 🌡 Temperature: 40°C

  5. Standard preparation

    Prepare calibration curve (5 points) from certified dye standards in range 0.5–50 µg/mL.

  6. HPLC analysis

    Inject 10–20 µL of sample onto C18 column. Gradient elution: 0→50% ACN in ammonium acetate buffer pH 7 in 25 min. Flow rate 1.0 mL/min.

    ⏱ Time: 30 min

  7. Detection and identification

    Record chromatogram at 427, 482, 520 and 630 nm. Identify dyes by retention time and UV-Vis absorption spectrum (DAD library).

  8. Quantitative determination

    Calculate dye concentrations from calibration curve. Include dilution factor and recovery.

  9. Quality control

    Analyze blank sample, spiked sample and CRM. Recovery 90–110%, repeatability RSD ≤ 5%.

    ⏱ Time: 30 min

  10. Report

    Prepare report with results in mg/kg, uncertainty and comparison with limits according to Regulation 1333/2008.

Required equipment and apparatus

EquipmentExampleIndicative price
HPLC system with DAD detectorAgilent 1260 Infinity II + DAD G7117C, Shimadzu Nexera + SPD-M40, Waters Alliance + 2998 PDA150 000–350 000 PLN
Chromatographic column C18Agilent Zorbax Eclipse Plus C18, Waters Symmetry C18, Phenomenex Kinetex C181 500–3 500 PLN
SPE system (solid phase extraction)Supelco Visiprep SPE manifold, Oasis HLB or polyamide cartridges2 000–5 000 PLN (manifold)
Ultrasonic bathBandelin Sonorex, Elma S, Branson 28003 000–8 000 PLN
Analytical balanceMettler Toledo MS205DU, RADWAG AS 220.R28 000–25 000 PLN

Reagents, media and consumables

ReagentCASDetails
Dye standards (certified CRM)Tartrazine E102, sunset yellow E110, azorubine E122, ponceau 4R E124, allura red E129, brilliant blue E133 — Sigma-Aldrich, Dr. Ehrenstorfer
Acetonitrile (HPLC grade)75-05-8Purity ≥ 99.9%, for mobile phase and extraction
Methanol (HPLC grade)67-56-1For extraction of dyes from food matrix
Ammonium acetate631-61-8For preparation of mobile phase buffer (pH 7.0, concentration 0.02 mol/L)
Ammonia 25%1336-21-6For alkalinization of extraction phase — facilitates dye dissolution
SPE cartridges (polyamide/C18)For purification of extracts from fat and protein matrices, 500 mg/6 mL

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN 12878:2014 (oraz metody wewnętrzne oparte na HPLC-DAD) — “Determination of synthetic food colorants by HPLC”.

How does this method work?

Synthetic dyes are extracted from a food sample with a mixture of water/methanol/ammonia, purified on SPE cartridges (polyamide or C18) and separated by liquid chromatography on a C18 column with gradient elution (mobile phase: ammonium acetate buffer pH 7 / acetonitrile). Detection using a DAD (diode array detector) at multiple wavelengths (427 nm, 482 nm, 520 nm, 630 nm) allows dye identification based on absorption spectrum and retention time.

What is the measuring range and accuracy?

Determination range: 0.5–500 mg/kg (depending on dye and matrix); Limit of detection (LOD): 0.1–0.5 mg/kg; Limit of quantification (LOQ): 0.5–1.0 mg/kg; Column: C18 (250 × 4.6 mm, 5 µm) or UHPLC C18 (100 × 2.1 mm, 1.8 µm).

How long does the test take?

The times given for the individual stages add up to approximately 1 h 45 min. The procedure comprises 10 steps.

What equipment is required?

HPLC system with DAD detector, Chromatographic column C18, SPE system (solid phase extraction), Ultrasonic bath, Analytical balance. Indicative cost of the core instrument (HPLC system with DAD detector): 150 000–350 000 PLN.

Where is this test used?

Official food control by Sanitary Inspection (SANEPID); Quality control of carbonated drinks, juices, syrups; Testing of sweets, candies, jellies for azo dyes; Analysis of spices and sauces (curcumin vs tartrazine); Control of children's food (EU restrictions — additional labeling); Testing of imported food for unauthorized dyes (Sudan I–IV); Monitoring of dyes in dietary supplements.

What safety precautions apply?

Acetonitrile — toxic (inhalation, skin contact), work under fume hood; Methanol — toxic, flammable (flash point 11°C) — fume hood, no open flame; Ammonia — corrosive, irritating to respiratory tract — fume hood, glasses, gloves; Dye standards — some azo dyes may be sensitizing, use gloves; HPLC waste (mobile phase) — collect in organic waste containers.

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 12878.

🔍 Find a laboratory performing this test