🍎 Sugar profile — HPLC

Food Chemistry PN-EN 12630

In short

Determination of glucose, fructose, sucrose, and sorbitol in food by HPLC with refractive index detector (RID). The test is performed according to PN-EN 12630:2001; Measurement range: 0.1–50 g/100 g. The procedure comprises 6 steps (about 1 h 45 min in total); it is used for: Labeling control — total sugar content (EU Reg. 1169/2011), Analysis of fruit juices — adulteration detection, Analysis of honey — sugar profile (authenticity).

At a glance

  • Standard: PN-EN 12630:2001
  • Category: Food Chemistry
  • Procedure steps: 6
  • Total time of stages: 1 h 45 min
  • Sugars determined: Glucose, fructose, sucrose, maltose, lactose, sorbitol
  • Measurement range: 0.1–50 g/100 g
  • Limit of detection: 0.05 g/100 g

Overview

Sugar profile is a key parameter in food analysis, used for labeling control (total sugar content), adulteration detection (e.g., addition of glucose-fructose syrup to juice), product characterization (glucose/fructose ratio), and quality control.

The HPLC method with refractive index detection (RID) allows simultaneous separation and determination of mono- and disaccharides and sugar alcohols (sorbitol) on an amino (NH₂) or ion-exchange column (Ca²⁺/Pb²⁺). Sample preparation includes water extraction, deproteination (Carrez I + II), and filtration.

Method principle

Sugars extracted from the sample with water solution and purified with Carrez reagents (protein precipitation) are separated on an HPLC column with aminopropyl phase (NH₂) or ion-exchange column (Ca²⁺). Refractive index detection (RID) measures the difference in refractive index between eluent and sample. Identification based on retention times, quantitative determination using a calibration curve.

Applications

Key parameters

ParameterValue
Sugars determinedGlucose, fructose, sucrose, maltose, lactose, sorbitol
Measurement range0.1–50 g/100 g
Limit of detection0.05 g/100 g
Precision (CV)1–3%
Analysis time15–25 min

Standard

Standard number
PN-EN 12630:2001
Title (PL)
Soki owocowe i warzywne — Oznaczanie glukozy, fruktozy, sorbitolu i sacharozy — Metoda HPLC
Title (EN)
Fruit and vegetable juices — Determination of glucose, fructose, sorbitol and sucrose — HPLC method

Step-by-step procedure

  1. Extraction

    Weigh 5 g sample into 100 mL flask. Add 50 mL warm water (60°C). Mix 30 min. Cool.

    ⏱ Time: 45 min

  2. Deproteination

    Add 2 mL Carrez I, mix. Add 2 mL Carrez II, mix. Fill to 100 mL with water.

    ⏱ Time: 5 min

  3. Filtration

    Filter through paper filter, then through 0.22 µm into HPLC vial.

    ⏱ Time: 10 min

  4. Calibration

    Prepare mixed sugar standard: 0.1–5 g/100 mL of each sugar.

    ⏱ Time: 20 min

  5. HPLC analysis

    NH₂ column, mobile phase: ACN/H₂O 75:25, flow 1.0 mL/min, RID detection, column temp. 30°C.

    ⏱ Time: 20 min

  6. Calculations

    Read concentrations from calibration curve. Total sugars = glucose + fructose + sucrose. Convert to g/100 g.

    ⏱ Time: 5 min

Required equipment and apparatus

EquipmentExampleIndicative price
HPLC system with RID detectorAgilent 1260 + RID, Shimadzu RID-20A, Waters 2414200 000–350 000 PLN
Aminopropyl column (NH₂)Shodex Asahipak NH2P-50 4E, Waters Sugar-Pak2 000–5 000 PLN
CentrifugeEppendorf 5804R10 000–30 000 PLN
Volumetric flask 100 mLClass A, calibrated20–50 PLN/pc

Reagents, media and consumables

ReagentCASDetails
D-Glucose (standard)50-99-7Purity ≥99.5%, certified CRM
D-Fructose (standard)57-48-7Purity ≥99%, certified CRM
Sucrose (standard)57-50-1Purity ≥99.5%, certified CRM
Carrez I reagentK₄[Fe(CN)₆] · 3H₂O, 150 g/L — for deproteination
Carrez II reagentZnSO₄ · 7H₂O, 300 g/L — for deproteination
Acetonitrile HPLC grade75-05-8Mobile phase: ACN/H₂O 75:25 (for NH₂ column)

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN 12630:2001 — “Fruit and vegetable juices — Determination of glucose, fructose, sorbitol and sucrose — HPLC method”.

How does this method work?

Sugars extracted from the sample with water solution and purified with Carrez reagents (protein precipitation) are separated on an HPLC column with aminopropyl phase (NH₂) or ion-exchange column (Ca²⁺). Refractive index detection (RID) measures the difference in refractive index between eluent and sample.

What is the measuring range and accuracy?

Sugars determined: Glucose, fructose, sucrose, maltose, lactose, sorbitol; Measurement range: 0.1–50 g/100 g; Limit of detection: 0.05 g/100 g; Precision (CV): 1–3%.

How long does the test take?

The times given for the individual stages add up to approximately 1 h 45 min. The procedure comprises 6 steps.

What equipment is required?

HPLC system with RID detector, Aminopropyl column (NH₂), Centrifuge, Volumetric flask 100 mL. Indicative cost of the core instrument (HPLC system with RID detector): 200 000–350 000 PLN.

Where is this test used?

Labeling control — total sugar content (EU Reg. 1169/2011); Analysis of fruit juices — adulteration detection; Analysis of honey — sugar profile (authenticity); Quality control of confectionery and beverages; Analysis of dietetic products (simple sugars); Quality control of alcoholic beverages (beer, wine).

What safety precautions apply?

Acetonitrile — toxic, flammable. Work in fume hood.; Carrez reagents — irritants, gloves and safety glasses; HPLC waste — collect separately.

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 12630.

🔍 Find a laboratory performing this test