⚗️ Phenol index in water — flow analysis (FIA/CFA) with 4-aminoantipyrine and potassium peroxodisulfate, in-line extraction into chloroform (470–475 nm)

Physicochemistry PN-EN ISO 14402

In short

The sample is introduced into the carrier stream and mixed with solutions of 4-aminoantipyrine (in a buffer of about pH 10. The test is performed according to PN-EN ISO 14402:2004 (ISO 14402:1999); Range: 0.01–1 mg/l (range II 0.01–0.1; range I 0.1–1 mg/l), a separate calibration with at least 5 standards; with detergents only above 0.1 mg/l. The procedure comprises 6 steps; it is used for: Industrial wastewater (coking plants, refineries, the chemical industry, tanneries) and municipal wastewater — the phenol index as a limited parameter (0.1 mg/l of volatile phenols), Surface water and groundwater, leachate from landfills — monitoring of phenols from 0.01 mg/l, High-throughput laboratories — automatic FIA/CFA analysis with in-line extraction into chloroform (without distillation).

At a glance

  • Standard: PN-EN ISO 14402:2004 (ISO 14402:1999)
  • Category: Physicochemistry
  • Procedure steps: 6
  • Range: 0.01–1 mg/l (range II 0.01–0.1; range I 0.1–1 mg/l), a separate calibration with at least 5 standards; with detergents only above 0.1 mg/l
  • Detection: 470–475 nm in the chloroform phase; FIA cell 10 mm; CFA cell 0.5–5 cm; peak heights
  • In-line reagents: R1: 0.5 g of 4-aminoantipyrine per 100 ml of buffer of pH 10.3 (NaHCO₃ 23 g, H₃BO₃ 27 g, KOH 35 g/l); R2: 5 g of K₂S₂O₈ per 100 ml at pH 11; both freshly prepared every day and degassed

Overview

ISO 14402:1999 specifies two methods for determining the phenol index in waters of various origin (groundwater, surface water, leachate, wastewater) at concentrations of 0.01–1 mg/l in the undiluted sample (the range can be adjusted by the working conditions): clause 3 — the phenol index without distillation, after extraction; clause 4 — the phenol index without extraction, after distillation. The 2019 regulation (Dz.U. 2019 poz. 1311 [Journal of Laws 2019, item 1311]) names PN-EN ISO 14402 (flow analysis, CFA and FIA) alongside PN-ISO 6439 (spectrophotometry) as the reference methods for the phenol index in wastewater, for which the limit for volatile phenols is 0.1 mg/l. In Poland it is in force as PN-EN ISO 14402:2004 (Polish version; it replaced PN-EN ISO 14402:2002).

Interferences: aromatic amines form condensation products with 4-aminoantipyrine (results too high); after the reagents have been added the sample has to reach a pH of 10.0–10.5 — strongly acid, strongly alkaline and buffered samples are brought beforehand to pH 5–7; turbidity does not interfere; with coloured samples it is checked whether the colour is extracted by chloroform, and a blank is measured without the reagents R1 and R2; detergents in wastewater interfere strongly (foam disturbs the distillation of the volatile phenols and the segmentation/separation of the phases) — where the detergent content is considerable, the standard applies only above 0.1 mg/l of phenol. The classical manual method (PWr textbook "Chemia wody" [Water chemistry], PN-ISO 6439) is based on the same reaction: after steam distillation the volatile phenols react at pH 10.0 ± 0.2 in the presence of potassium ferricyanide with 4-aminoantipyrine, the product being extracted with chloroform and measured at 510 nm (0.002–5.0 mg/dm³).

Method principle

The sample is introduced into a continuously flowing carrier stream and mixed with the likewise continuously flowing solutions of 4-aminoantipyrine and potassium peroxodisulfate. The phenolic compounds are oxidised by the peroxodisulfate, and the quinones formed react with the 4-aminoantipyrine, giving coloured condensation products which, in a flow extraction unit, pass from the aqueous phase into chloroform; the chloroform phase is separated by a phase separator (e.g. a hydrophobic semi-permeable PTFE membrane, thickness 150–200 µm, pores 0.5–2 µm, porosity 75 %), and the absorbance of the organic phase is measured by a flow spectrometer at 470–475 nm. Reagents (of analytical grade, the solutions degassed, best under reduced pressure): grade 1 water, KOH, NaHCO₃, 4-aminoantipyrine, K₂S₂O₈, phenol (colourless crystals), boric acid, ethanol 96 %, 2-propanol, H₂SO₄ 1.84 g/ml, HCl 50 %, KOH 1.0 mol/l; buffer (23 g of NaHCO₃ + 27 g of H₃BO₃ + 35 g of KOH per 1000 ml; about pH 10.3; 1 month); carrier C — degassed water; R1 — 0.5 g of 4-aminoantipyrine in 100 ml of the buffer (freshly prepared every day, degassed); R2 — 5 g of K₂S₂O₈ in about 90 ml of water, pH 11 (KOH 1 mol/l), made up to 100 ml (every day); Org — degassed chloroform (membrane or 3 min of ultrasound); phenol stock solution 1000 mg/l (1.000 g/l; 1 month at 2–5 °C), standards of 10 mg/l (a week) and 1 mg/l; calibration solutions (at least 5 per range, freshly prepared every day): range I 0.1–1 mg/l (1, 3, 5, 6, 8, 10 ml of the 10 mg/l standard made up to 100 ml), range II 0.01–0.1 mg/l (the same volumes of the 1 mg/l standard). FIA apparatus: reagent reservoirs, a low-pulsation pump, a displacement bottle for the chloroform, an injector (600 µl for the range 0.01–0.1 mg/l, 200 µl for 0.1–1.0 mg/l), an extraction cell with a segmenter and a phase separator, tubing and coils of 0.5–0.8 mm (reaction coils of 60 and 80 cm, extraction coil 160 cm × 0.7 mm), a detector with a 10 mm cell at 470–475 nm, a recorder (peak heights), an autosampler; CFA: an autosampler, a pump with inert tubing, a manifold with air segmentation (e.g. internal diameter 2 mm; coils of 40, 40 and 150 cm), a phase separator, a detector with a cell of 0.5–5 cm at 470–475 nm. In addition: flasks of 100 and 1000 ml, pipettes of 1–10 ml, membrane filters 0.45 µm, a pH meter. Samples: glass or PTFE vessels, rinsed before use with sulfuric acid of about pH 2; analysis immediately after sampling, or acidification to about pH 2 (H₂SO₄ or HCl), storage in the dark at 2–5 °C and analysis within 24 h (exceptionally, after acidification and pressure filtration, for up to 2 weeks after verification); filtration where there is a risk of blocking. Preparation for the measurement: pass C, R1, R2 and Org through until the baseline is stable, then zero it; a poor signal-to-noise ratio usually results from a damaged separator membrane or traces of water on the walls of the cell (rinse with ethanol or 2-propanol). Check of the system: in range II a standard of 0.05 mg/l has to give an absorbance of at least 0.01 cm⁻¹; reagent blank: water instead of R1 and R2 — a drop of absorbance above 0.05 cm⁻¹ means self-condensation products (repeat the preparation of the solutions). Calibration separately for each range (FIA: 200 µl for I, 600 µl for II; CFA: choice of the cell length and of the flow rate for the highest response).

Applications

Key parameters

ParameterValue
Range0.01–1 mg/l (range II 0.01–0.1; range I 0.1–1 mg/l), a separate calibration with at least 5 standards; with detergents only above 0.1 mg/l
Detection470–475 nm in the chloroform phase; FIA cell 10 mm; CFA cell 0.5–5 cm; peak heights
In-line reagentsR1: 0.5 g of 4-aminoantipyrine per 100 ml of buffer of pH 10.3 (NaHCO₃ 23 g, H₃BO₃ 27 g, KOH 35 g/l); R2: 5 g of K₂S₂O₈ per 100 ml at pH 11; both freshly prepared every day and degassed
Extractiondegassed chloroform; phase separator — a PTFE membrane 150–200 µm, pores 0.5–2 µm, porosity 75 %
Injection volume (FIA)600 µl (0.01–0.1 mg/l), 200 µl (0.1–1.0 mg/l); coils of 60/80 cm × 0.5 mm, extraction coil 160 cm × 0.7 mm
Controlstandard of 0.05 mg/l → A ≥ 0.01 cm⁻¹; reagent blank: drop ≤ 0.05 cm⁻¹; pH of the sample after the reagents 10.0–10.5; samples: pH 2, 2–5 °C, 24 h

Standard

Standard number
PN-EN ISO 14402:2004 (ISO 14402:1999)
Title (PL)
Jakość wody — Oznaczanie indeksu fenolowego za pomocą analizy przepływowej (FIA i CFA)
Title (EN)
Water quality — Determination of phenol index by flow analysis (FIA and CFA)

Step-by-step procedure

  1. Sample

    Rinse the glass or PTFE vessels with sulfuric acid of about pH 2. Analyse at once or acidify to about pH 2 (H₂SO₄ or HCl), store in the dark at 2–5 °C and examine within 24 h (exceptionally for up to 2 weeks after acidification and pressure filtration, once verified); filter where there is a risk of blocking the tubing; bring strongly acid, strongly alkaline or buffered samples to pH 5–7.

  2. Reagents

    Prepare fresh R1 (0.5 g of 4-aminoantipyrine in 100 ml of the buffer) and R2 (5 g of K₂S₂O₈ in 100 ml, pH 11) every day, degas them (membrane), degas the chloroform (membrane or 3 min of ultrasound) and the carrier water (under reduced pressure); prepare at least 5 calibration solutions of the chosen range (I: 0.1–1.0 mg/l; II: 0.01–0.1 mg/l).

  3. Starting up the system

    Pass C, R1, R2 and Org through the system until the baseline is stable and zero it; check the signal-to-noise ratio (a damaged separator membrane or water on the walls of the cell — rinse with ethanol/2-propanol).

  4. Check of the system and of the blank

    In range II a standard of 0.05 mg/l has to give an absorbance of at least 0.01 cm⁻¹ (otherwise the system is not suitable). Instead of R1 and R2 pass water through until the signal is constant — a drop of absorbance above 0.05 cm⁻¹ means self-condensation products: prepare the solutions again and repeat the checks; then restore R1 and R2.

  5. Calibration

    Calibrate separately for each range: FIA — an injection of 200 µl (range I) or 600 µl (range II); CFA — choose the cell length and the flow rate for the highest response of the highest standard; record the peak heights.

  6. Measurement and result

    Measure the samples under the same conditions; for coloured samples check whether the colour is extracted by chloroform and measure a blank without R1 and R2, taking the difference into account in the calculations; read the phenol index (mg/l as phenol) from the calibration; where detergents are present report only results above 0.1 mg/l; record the variant (extraction/distillation), the preservation and any deviations.

Required equipment and apparatus

EquipmentExampleIndicative price
FIA analyser (low-pulsation pump, injector of 200/600 µl) or CFA analyser (autosampler, manifold with air segmentation)tubing of 0.5–0.8 mm (FIA) or about 1–2 mm (CFA) of inert plastics—
Displacement bottle for feeding the chloroformthe Org stream—
Extraction cell with a phase segmenter and a PTFE membrane separatorin-line extraction of the condensation products into chloroform—
Flow spectrometer at 470–475 nm (10 mm cell in FIA; 0.5–5 cm in CFA) and a recordermeasurement of peak heights—
Volumetric flasks of 100 and 1000 ml, pipettes of 1–10 ml, membrane filters 0.45 µm, pH meterstandards and preparation of the samples—
Ultrasonic bath or a membrane degassing setdegassing of the chloroform and of the reagents—

Reagents, media and consumables

ReagentCASDetails
4-aminoantipyrine (solution of 0.5 g per 100 ml in the borate-hydrogen carbonate buffer of pH 10.3)—reagent coupling with the quinones (R1)
Potassium peroxodisulfate 5 g per 100 ml, pH 11—oxidant for the phenols (R2)
Buffer: NaHCO₃ 23 g + H₃BO₃ 27 g + KOH 35 g per 1000 ml—pH of about 10.3 for the reaction
Degassed chloroform—extraction phase (Org)
Phenol (colourless crystals) — 1000 mg/l, standards of 10 and 1 mg/l—calibration of ranges I and II
Sulfuric acid / hydrochloric acid 50 %, KOH 1 mol/l, ethanol 96 %, 2-propanol—preservation of the samples (pH 2), pH adjustment, rinsing of the cell

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN ISO 14402:2004 (ISO 14402:1999) — “Water quality — Determination of phenol index by flow analysis (FIA and CFA)”.

How does this method work?

The sample is introduced into a continuously flowing carrier stream and mixed with the likewise continuously flowing solutions of 4-aminoantipyrine and potassium peroxodisulfate. The phenolic compounds are oxidised by the peroxodisulfate, and the quinones formed react with the 4-aminoantipyrine, giving coloured condensation products which, in a flow extraction unit, pass from the aqueous phase into chloroform; the chloroform phase is separated by a phase separator (e.

What is the measuring range and accuracy?

Range: 0.01–1 mg/l (range II 0.01–0.1; range I 0.1–1 mg/l), a separate calibration with at least 5 standards; with detergents only above 0.1 mg/l; Detection: 470–475 nm in the chloroform phase; FIA cell 10 mm; CFA cell 0.5–5 cm; peak heights; In-line reagents: R1: 0.5 g of 4-aminoantipyrine per 100 ml of buffer of pH 10.3 (NaHCO₃ 23 g, H₃BO₃ 27 g, KOH 35 g/l); R2: 5 g of K₂S₂O₈ per 100 ml at pH 11; both freshly prepared every day and degassed; Extraction: degassed chloroform; phase separator — a PTFE membrane 150–200 µm, pores 0.5–2 µm, porosity 75 %.

How long does the test take?

The procedure comprises 6 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

FIA analyser (low-pulsation pump, injector of 200/600 µl) or CFA analyser (autosampler, manifold with air segmentation), Displacement bottle for feeding the chloroform, Extraction cell with a phase segmenter and a PTFE membrane separator, Flow spectrometer at 470–475 nm (10 mm cell in FIA; 0.5–5 cm in CFA) and a recorder, Volumetric flasks of 100 and 1000 ml, pipettes of 1–10 ml, membrane filters 0.45 µm, pH meter, Ultrasonic bath or a membrane degassing set.

Where is this test used?

Industrial wastewater (coking plants, refineries, the chemical industry, tanneries) and municipal wastewater — the phenol index as a limited parameter (0.1 mg/l of volatile phenols); Surface water and groundwater, leachate from landfills — monitoring of phenols from 0.01 mg/l; High-throughput laboratories — automatic FIA/CFA analysis with in-line extraction into chloroform (without distillation); Turbid samples (which do not interfere) and coloured samples (with a blank measured without the reagents); The variant with distillation (clause 4 of the standard) for samples with non-extractable substances or where volatile phenols are required.

What safety precautions apply?

Phenol is toxic and is readily absorbed through the skin — gloves, goggles; chloroform is toxic and carcinogenic — work in a fume hood, chloroform and phenol waste to be disposed of; Potassium peroxodisulfate — an oxidant; potassium hydroxide and concentrated sulfuric acid — corrosive; Ethanol and 2-propanol — flammable.