⚗️ Sludge volume after 30 min of settling and the sludge volume index (Mohlman index) — a 1000 ml cylinder, determination of the dry matter and the quotient of the two quantities
In short
The sludge suspension is poured with a scoop into a 1000 ml measuring cylinder and after 30 min without shaking the sludge volume is read at the sludge–water boundary. The test is performed according to PN-EN 14702-1:2008; Resulting quantities: the proportion of sludge volume after settling V_s in ml/l (given rounded to 10 ml/l; the standard distinguishes the sludge volume in ml from the proportion of sludge volume in ml/l — the symbol V_s denotes the latter) and the sludge volume index I_sv in ml/g (given rounded to 1 ml/g). The procedure comprises 11 steps (about 1 h in total); it is used for: Routine operational control of activated-sludge tanks and secondary settlers in municipal wastewater treatment plants, Early detection of sludge bulking caused by the development of filamentous bacteria, before the sludge is carried out of the settler, Assessment of the working of treatment plants for industrial wastewater of a character close to municipal wastewater (within the meaning of Directive 91/271/EEC).
At a glance
- Standard: PN-EN 14702-1:2008
- Category: Physicochemistry
- Procedure steps: 11
- Total time of stages: 1 h
- Resulting quantities: the proportion of sludge volume after settling V_s in ml/l (given rounded to 10 ml/l; the standard distinguishes the sludge volume in ml from the proportion of sludge volume in ml/l — the symbol V_s denotes the latter) and the sludge volume index I_sv in ml/g (given rounded to 1 ml/g)
- Formula: I_sv = V_s / ρ_T, where V_s is the proportion of sludge volume after 30 min of settling in ml/l, taken as the mean of at least two measurements, and ρ_T is the dry matter concentration in the sludge in g/l
- Settling time: 30 min without shaking, in one place
Overview
This is the simplest and the most frequently carried out determination in the operation of an activated-sludge plant: a cylinder, a clock and a balance. The introduction to the standard puts the matter plainly — determining the settleability by establishing the proportion of sludge volume is an operational parameter and makes it possible to assess (only within certain limits, admittedly) the dry matter content in the mixture of sludge and water; the value serves for calculating the settling velocity of the sludge, for calculating the sludge volume index and for assessing the working of settlers. The scope covers sludge suspensions originating from the management of rainwater, from municipal wastewater collecting systems, from municipal wastewater treatment plants, from treatment plants for industrial wastewater similar to municipal wastewater (within the meaning of Directive 91/271/EEC) and from waterworks; the method also applies to sludge suspensions of other origin.
What this indicator really says. The sludge volume index (in Polish literature: indeks osadu, the Mohlman index, IO; in English-language literature SVI) is the quotient of the sludge volume and the dry matter concentration in the sludge. The lower its value, the more favourable the settling properties of the sludge. The laboratory instruction of Wrocław University of Science and Technology states that activated sludge with good settling properties has an IO within the range 50–100 cm³/g DM, and that higher values point to so-called sludge bulking, caused most often by filamentous bacteria. A survey of technological parameters in a paper in "Gaz, Woda i Technika Sanitarna" gives 80–120 ml/g DM as the optimum value and explains the mechanism: high IO values follow from the intensive development of filamentous bacteria caused by unfavourable operating conditions, and a large rise in this parameter may lead to the aerobic stabilisation of the sludge, whereby the flocs become smaller and settle worse. The practical sense is this: a rising index is a warning signal of a deterioration of settling in the secondary settler — which is why the determination is carried out routinely and interpreted in a time series, not occasionally.
Which sheet of the series is which document. The series PN-EN 14702 „Charakterystyka osadów ściekowych — Właściwości sedymentacyjne" [Characterisation of sludges — Settling properties] contains no sheet with requirements — all three parts are standards of test methods and differ in what they measure. Part 1 (this entry) describes settleability: free settling for 30 min in a 1000 ml cylinder, the proportion of sludge volume and the sludge volume index. Part 2 describes thickenability: the same 30 min, but in a cylinder of 10 cm diameter and 50 cm measuring height (nominal volume 3500 ml) with a slow stirrer at 1 rev/min — there one measures the stirred sludge volume and the stirred sludge volume index, and the standard notes that the settling properties depend on the concentration of the suspension, so that comparing sludges of different concentration is admissible if a reference concentration is adopted, for which 3.5 g/l has been taken. Part 3, considerably younger, concerns the determination of the zone settling velocity (ZSV). For the ordinary "sludge index" of treatment-plant operation only Part 1 is the right one.
Status and editions. The current edition is PN-EN 14702-1:2008 — the Polish-language version, published on 2 December 2008, 15 pages, price group H, managed by KT 119 for Water Quality — Basic Problems, ICS 13.030.20. It implements EN 14702-1:2006 [IDT] and supersedes PN-EN 14702-1:2006 — the English-language version. The PKN catalogue card for this edition has neither a "Date of withdrawal" field, nor a "Superseded by" field, nor a corrigendum (Ap). Only this one language version has a current card — the English version exists solely as the withdrawn edition of 2006 — so the trap of "the corrigendum hangs only on one version" does not occur here. The other items of the series in the PKN list are: PN-EN 14702-2:2008 — the Polish version (implementing EN 14702-2:2006 [IDT]) and PN-EN 14702-3:2019-04 — the English version (implementing EN 14702-3:2019 [IDT]); the English versions of Parts 1 and 2 of 2006 bear the annotation "Withdrawn and superseded by" the corresponding Polish versions of 2008. It is worth noting a small point which is sometimes confused when the title is copied out: the title of the withdrawn English version of 2006 read „Oznaczanie zdolności opadania (Oznaczanie stosunku objętości osadu do indeksu objętościowego osadu)", whereas the binding Polish version of 2008 has the title „Oznaczanie zdolności opadania osadu (Oznaczanie objętości względnej oraz indeksu objętościowego osadu)" — the second wording is the correct one. EN 14702-1:2006 was prepared by technical committee CEN/TC 308 "Characterisation of sludges", whose secretariat is held by AFNOR; the standard was approved by CEN on 3 February 2006, and conflicting national standards were to be withdrawn at the latest by September 2006.
What this description does not contain. We have not seen a Polish copy of PN-EN 14702-1:2008 — only the title, the scope and the publication data come from the PKN catalogue card. The substantive content is given from the preview of SIST EN 14702-1:2006 — the Slovenian implementation of the same European standard, identical with EN 14702-1:2006 (and therefore with the text which PN-EN 14702-1:2008 implements as IDT) — covering clauses 1–10 in full. We have not seen Annex A (informative) with the results of the validation trials or the bibliography, so we give only those precision data which stand in clause 9 of the main text. The threshold values for interpreting the index (50–100 or 80–120 ml/g) do not come from the standard — the standard gives no limit values at all — but from technological literature, and are marked as such in the text.
Method principle
The sludge volume after settling and the sludge volume index are determined by a 30-minute settling of the sludge suspension. A representative sample of the suspension is taken with a scoop and poured at once into a measuring cylinder up to the 1000 ml mark; after 30 minutes standing in one place, without shaking, the sludge volume is read at the level of the sludge surface, that is at the sludge–water boundary. The sludge volume divided by the volume of the initial sample used gives the proportion of sludge volume in ml/l, and the same volume divided by the dry matter concentration in the sludge, expressed in g/l, gives the sludge volume index in ml/g. Settling takes place solely under the force of gravity, without stirring — in this the method differs from the determination of thickenability according to Part 2 of the standard, where the cylinder is stirred with a slow stirrer.
Applications
- Routine operational control of activated-sludge tanks and secondary settlers in municipal wastewater treatment plants
- Early detection of sludge bulking caused by the development of filamentous bacteria, before the sludge is carried out of the settler
- Assessment of the working of treatment plants for industrial wastewater of a character close to municipal wastewater (within the meaning of Directive 91/271/EEC)
- Sludges from the management of rainwater, from sewer systems and from waterworks
- Choice of the sludge recirculation ratio and estimation of the settling velocity of the sludge for technological calculations
- Comparative examination of sludges before a change of process parameters (sludge age, loading, aeration)
Key parameters
| Parameter | Value |
|---|---|
| Resulting quantities | the proportion of sludge volume after settling V_s in ml/l (given rounded to 10 ml/l; the standard distinguishes the sludge volume in ml from the proportion of sludge volume in ml/l — the symbol V_s denotes the latter) and the sludge volume index I_sv in ml/g (given rounded to 1 ml/g) |
| Formula | I_sv = V_s / ρ_T, where V_s is the proportion of sludge volume after 30 min of settling in ml/l, taken as the mean of at least two measurements, and ρ_T is the dry matter concentration in the sludge in g/l |
| Settling time | 30 min without shaking, in one place |
| Measuring vessel | a measuring cylinder of 1000 ml nominal volume, of glass or transparent plastic, 60–70 mm in diameter; when the sludge volume after 30 min is less than 50 ml, an Imhoff cone of 1 l capacity may be used |
| Dilution threshold | the determination is repeated if the sludge volume exceeds 250 ml/l; it is diluted with the supernatant of the suspension or with water from the settler outlet in a volume ratio of 1:1, 1:2 or 1:3, and the reading is multiplied by 2, 3 or 4. The test report states the dilution at which the value first fell below 250 ml/l |
| Number of replicates | the determination is carried out in duplicate; the mean of at least two measurements enters the formula |
| Determination of the dry matter | when the concentration of dissolved matter is less than 10 % of the total matter — according to EN 12880; when greater than 10 % — according to EN 872 |
| Precision (clause 9 of the standard) | standard deviation of repeatability from 0.066 ml/g (0.2 %) for digested sludge through 0.287 ml/g (1.2 %) for waterworks sludge to 4.370 ml/g (3.0 %) for activated sludge; mean value 1.574 ml/g (2.2 %), minimum precision 3.0 %. Standard deviation of reproducibility from 0.131 ml/g (0.3 %) through 0.521 ml/g (2.2 %) to 7.304 ml/g (5.1 %); mean value 2.652 ml/g (3.8 %), minimum precision 5.1 % |
| Interpretative values (from outside the standard) | technological literature: sludge with good settling properties 50–100 cm³/g DM (the Wrocław University of Science and Technology laboratory instruction), optimum value 80–120 ml/g DM (a paper in "Gaz, Woda i Technika Sanitarna"); higher values point to sludge bulking. The standard itself gives no limit values |
Standard
- Standard number
- PN-EN 14702-1:2008
- Title (PL)
- Charakterystyka osadów ściekowych — Właściwości sedymentacyjne — Część 1: Oznaczanie zdolności opadania osadu (Oznaczanie objętości względnej oraz indeksu objętościowego osadu)
- Title (EN)
- Characterisation of sludges — Settling properties — Part 1: Determination of settleability (Determination of the proportion of sludge volume and sludge volume index)
Step-by-step procedure
-
Checking the conditions before taking the sample
Establish the temperature of the sample and the ambient temperature. When the difference is greater than 5 °C the standard recommends standing the settling cylinder with the sample in a bucket filled with the liquid of the sample — otherwise convection and the release of gas bubbles will disturb the settling.
-
Taking the sample with a scoop
Take a representative sample of the sludge suspension with a scoop holding 1 l to the brim and pour it at once into a measuring cylinder up to the 1000 ml mark. The use of a scoop of that capacity has a definite purpose: it prevents the sludge from settling in the scoop itself before it is poured out. The mixture of sludge and water must not be shaken too vigorously, so as not to change the course of settling.
-
Settling for 30 minutes
Leave the cylinder in one place for 30 min, without shaking. Do not move it, do not stir it, do not take samples during that time — the process is disturbed both by the walls of the vessel and by the mutual interaction of the flocs, especially when the proportion of sludge volume is large.
-
Reading the sludge volume
After 30 min read the sludge volume at the level of the sludge surface, that is at the boundary between the sludge and the supernatant water. The reading is unambiguous only when the boundary is distinct — its blurring is in itself information about the state of the sludge.
-
Deciding on dilution
If the sludge volume exceeds 250 ml/l, the determination has to be repeated on a diluted sample. For the dilution use the supernatant of the sludge suspension or water from the settler outlet, in a volume ratio of 1:1, 1:2 or 1:3. Homogenise the diluted sample gently — by turning the closed cylinder upside down twice.
-
Repeating the settling for the diluted sample
Repeat the 30 min settling and the reading. In the calculations multiply the sludge volume read off by the dilution factor 2, 3 or 4 — corresponding to the dilution 1:1, 1:2 and 1:3. The test report states the dilution at which the value first fell below 250 ml/l.
-
Duplicate
The determination is carried out in duplicate. The sludge volume which enters the formula for the sludge volume index is the mean of at least two measurements.
-
Determining the dry matter concentration
If the concentration of total dissolved matter is less than 10 % of the total matter — determine the mass concentration of dry matter in the sludge (g/l) according to EN 12880. If it is greater than 10 % — according to EN 872. The standard also notes that when the content of dissolved matter is small and negligible in relation to the dry matter content, the total matter content is to be used in the calculations.
-
Calculating the proportion of sludge volume and the index
The proportion of sludge volume V_s in ml/l is obtained as the sludge volume after settling divided by the volume of the initial sample used. The sludge volume index is calculated from the formula I_sv = V_s / ρ_T, where ρ_T is the dry matter concentration in g/l. The proportion of sludge volume is given rounded to 10 ml/l, the index — to 1 ml/g. The standard gives examples of the wording: „Objętość względna osadu po sedymentacji 180 ml/l" [proportion of sludge volume after settling 180 ml/l] for the original sample, „510 ml/l (170 ml po trzykrotnym rozcieńczeniu)" [510 ml/l (170 ml after a threefold dilution)] for a diluted sample, and „Indeks objętościowy osadu 145 ml/g" [sludge volume index 145 ml/g].
-
Test report
In the test report give: the reference to the standard, all the information needed for the full identification of the sludge sample, the details of the preparation of the sample (e.g. the dilution), the results calculated according to clause 8, and any details not specified in the standard or optional and any other factor which may have influenced the results.
-
Use of the result in operation
Interpret the result in a time series, not singly. Technological literature takes it that sludge with good settling properties has an index of 50–100 cm³/g DM (the Wrocław University of Science and Technology laboratory instruction), and that the optimum values lie within the range 80–120 ml/g DM; higher values point to sludge bulking, caused most often by filamentous bacteria developing under unfavourable operating conditions. The standard itself gives no thresholds — the limit values are a technological finding, not a normative one.
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Measuring cylinder 1000 ml | nominal volume 1000 ml, glass or transparent plastic, diameter 60–70 mm; when the sludge volume after 30 min is below 50 ml, an Imhoff cone of 1 l capacity is admissible | — |
| Scoop of 1 l nominal capacity | the standard requires a scoop holding 1 l to the brim — the sample is poured out of it at once, which prevents settling already in the scoop | — |
| Clock or timer | the settling time of 30 min is measured from the moment the cylinder is filled | — |
| Equipment for determining the dry matter | analytical balance, drying oven and weighing dishes (determination according to EN 12880) or a set for filtration through glass fibre filters (according to EN 872) — without the dry matter concentration the index cannot be calculated | — |
| Vessel for a temperature-equalising bath | a bucket filled with the liquid of the sample; the standard recommends standing the cylinder in it when the difference between the temperature of the sample and the ambient temperature exceeds 5 °C | — |
| Water for dilution | the supernatant of the sludge suspension or water from the settler outlet — not tap water and not distilled water | — |
Health and safety (OHS)
- Activated sludge suspension and raw sludge are infectious material — work in gloves, avoid aerosol when pouring from the scoop, do not touch the face, and disinfect hands and vessels after work.
- Taking a sample from an aeration tank or from a settler channel takes place over an open basin — use a scoop on a handle, do not lean over the edge, and apply fall protection in accordance with the instructions for the works.
- In channels and closed chambers hydrogen sulfide and methane may accumulate and the oxygen concentration is sometimes lowered — such places are entered only under a confined-space working procedure, with measurement of the atmosphere and a standby person.
- Return the sludge from the sample to the process after the determination or dispose of it as waste — do not pour it into the laboratory sink, where it blocks traps and colonises the installation.
- A 1000 ml cylinder filled with suspension is heavy and unstable — stand it on the bench away from the edge and from thoroughfares, because knocking it over interrupts the determination and spills infectious material.
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN 14702-1:2008 — “Characterisation of sludges — Settling properties — Part 1: Determination of settleability (Determination of the proportion of sludge volume and sludge volume index)”.
How does this method work?
The sludge volume after settling and the sludge volume index are determined by a 30-minute settling of the sludge suspension. A representative sample of the suspension is taken with a scoop and poured at once into a measuring cylinder up to the 1000 ml mark; after 30 minutes standing in one place, without shaking, the sludge volume is read at the level of the sludge surface, that is at the sludge–water boundary.
What is the measuring range and accuracy?
Resulting quantities: the proportion of sludge volume after settling V_s in ml/l (given rounded to 10 ml/l; the standard distinguishes the sludge volume in ml from the proportion of sludge volume in ml/l — the symbol V_s denotes the latter) and the sludge volume index I_sv in ml/g (given rounded to 1 ml/g); Formula: I_sv = V_s / ρ_T, where V_s is the proportion of sludge volume after 30 min of settling in ml/l, taken as the mean of at least two measurements, and ρ_T is the dry matter concentration in the sludge in g/l; Settling time: 30 min without shaking, in one place; Measuring vessel: a measuring cylinder of 1000 ml nominal volume, of glass or transparent plastic, 60–70 mm in diameter; when the sludge volume after 30 min is less than 50 ml, an Imhoff cone of 1 l capacity may be used.
How long does the test take?
The times given for the individual stages add up to approximately 1 h. The procedure comprises 11 steps.
What equipment is required?
Measuring cylinder 1000 ml, Scoop of 1 l nominal capacity, Clock or timer, Equipment for determining the dry matter, Vessel for a temperature-equalising bath, Water for dilution.
Where is this test used?
Routine operational control of activated-sludge tanks and secondary settlers in municipal wastewater treatment plants; Early detection of sludge bulking caused by the development of filamentous bacteria, before the sludge is carried out of the settler; Assessment of the working of treatment plants for industrial wastewater of a character close to municipal wastewater (within the meaning of Directive 91/271/EEC); Sludges from the management of rainwater, from sewer systems and from waterworks; Choice of the sludge recirculation ratio and estimation of the settling velocity of the sludge for technological calculations; Comparative examination of sludges before a change of process parameters (sludge age, loading, aeration).
What safety precautions apply?
Activated sludge suspension and raw sludge are infectious material — work in gloves, avoid aerosol when pouring from the scoop, do not touch the face, and disinfect hands and vessels after work.; Taking a sample from an aeration tank or from a settler channel takes place over an open basin — use a scoop on a handle, do not lean over the edge, and apply fall protection in accordance with the instructions for the works.; In channels and closed chambers hydrogen sulfide and methane may accumulate and the oxygen concentration is sometimes lowered — such places are entered only under a confined-space working procedure, with measurement of the atmosphere and a standby person.; Return the sludge from the sample to the process after the determination or dispose of it as waste — do not pour it into the laboratory sink, where it blocks traps and colonises the installation.; A 1000 ml cylinder filled with suspension is heavy and unstable — stand it on the bench away from the edge and from thoroughfares, because knocking it over interrupts the determination and spills infectious material.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 14702-1.