🍎 Preservatives (sorbates, benzoates) — HPLC

Food Chemistry PN-EN 12856

In short

Determination of sorbic acid and benzoic acid (preservatives E200, E210) in food by HPLC-UV. The test is performed according to PN-EN 12856:2001; Measurement range: 5–5000 mg/kg. The procedure comprises 5 steps (about 1 h 5 min in total); it is used for: Control of preservative use in food (EU Reg. 1333/2008), Analysis of soft drinks and juices, Analysis of fruit products (jams, marmalades).

At a glance

  • Standard: PN-EN 12856:2001
  • Category: Food Chemistry
  • Procedure steps: 5
  • Total time of stages: 1 h 5 min
  • Substances determined: Sorbic acid (E200), benzoic acid (E210)
  • Measurement range: 5–5000 mg/kg
  • Limit of detection: 2 mg/kg

Overview

Sorbic acid (E200) and benzoic acid (E210) and their salts are the most commonly used preservatives in food. Their permitted concentrations are regulated by EU Regulation 1333/2008: sorbic acid 150–2000 mg/kg (depending on product category), benzoic acid 150–2000 mg/kg.

The HPLC-UV method involves extraction of preservatives from food sample, purification, separation on a C18 column, and UV detection: sorbic acid 255 nm, benzoic acid 228 nm. The method is simple, fast, and selective.

Method principle

The sample is homogenized with water, ultrasonically extracted, and deproteinated with Carrez reagents. After filtration (0.22 µm), the solution is analyzed by HPLC on a C18 column with buffered mobile phase (ammonium acetate pH 4.2 + methanol). UV detection at 228 nm (benzoic acid) and 255 nm (sorbic acid). Quantitative determination using a calibration curve.

Applications

Key parameters

ParameterValue
Substances determinedSorbic acid (E200), benzoic acid (E210)
Measurement range5–5000 mg/kg
Limit of detection2 mg/kg
Wavelength228 nm (benzoic), 255 nm (sorbic)
Precision (CV)2–5%

Standard

Standard number
PN-EN 12856:2001
Title (PL)
Artykuły żywnościowe — Oznaczanie kwasu sorbowego i benzoesowego metodą HPLC
Title (EN)
Foodstuffs — Determination of acesulfame-K, aspartame and saccharin — HPLC method

Step-by-step procedure

  1. Extraction

    Weigh 5 g sample, add 40 mL water, ultrasonic 15 min. Deproteinate with Carrez I+II. Fill to 50 mL.

    ⏱ Time: 25 min

  2. Filtration

    Filter through 0.22 µm PVDF into vial.

    ⏱ Time: 5 min

  3. Calibration

    Standards: 5, 10, 50, 100, 500 mg/L mixture of sorbic and benzoic acid.

    ⏱ Time: 15 min

  4. HPLC analysis

    C18, 20 mM ammonium acetate pH 4.2 / MeOH gradient, 1 mL/min, UV 228+255 nm.

    ⏱ Time: 15 min

  5. Calculations

    Concentration from calibration curve, conversion to mg/kg product.

    ⏱ Time: 5 min

Required equipment and apparatus

EquipmentExampleIndicative price
HPLC system with UV/DAD detectorAgilent 1260, Shimadzu Nexera, Waters Alliance200 000–400 000 PLN
C18 columnAgilent Zorbax Eclipse Plus C18, 150×4.6 mm, 5 µm1 500–3 500 PLN
Ultrasonic bathBandelin Sonorex RK 5103 000–8 000 PLN
Syringe filters 0.22 µmPVDF or nylon, ø 13 mm1–3 PLN/pc

Reagents, media and consumables

ReagentCASDetails
Sorbic acid (standard)110-44-1Purity ≥99%, CRM
Benzoic acid (standard)65-85-0Purity ≥99.5%, CRM
Ammonium acetate631-61-8Mobile phase: 20 mmol/L, pH 4.2
Methanol HPLC grade67-56-1Mobile phase component (gradient 10–60%)

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN 12856:2001 — “Foodstuffs — Determination of acesulfame-K, aspartame and saccharin — HPLC method”.

How does this method work?

The sample is homogenized with water, ultrasonically extracted, and deproteinated with Carrez reagents. After filtration (0.

What is the measuring range and accuracy?

Substances determined: Sorbic acid (E200), benzoic acid (E210); Measurement range: 5–5000 mg/kg; Limit of detection: 2 mg/kg; Wavelength: 228 nm (benzoic), 255 nm (sorbic).

How long does the test take?

The times given for the individual stages add up to approximately 1 h 5 min. The procedure comprises 5 steps.

What equipment is required?

HPLC system with UV/DAD detector, C18 column, Ultrasonic bath, Syringe filters 0.22 µm. Indicative cost of the core instrument (HPLC system with UV/DAD detector): 200 000–400 000 PLN.

Where is this test used?

Control of preservative use in food (EU Reg. 1333/2008); Analysis of soft drinks and juices; Analysis of fruit products (jams, marmalades); Quality control of dairy products (cheeses, yogurts); Analysis of sauces, dressings, and marinades; Detection of undeclared preservatives.

What safety precautions apply?

Methanol — toxic, flammable. Fume hood mandatory.; Sorbic acid — irritant, nitrile gloves; HPLC waste — organic waste container.

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 12856.

🔍 Find a laboratory performing this test