🌿 Manganese and its compounds in workplace air — flame AAS
In short
Parallel sampling of the inhalable and of the respirable fraction of the aerosol of manganese on membrane filters, mineralisation with concentrated nitric acid, solutions in dilute HNO₃ with lanthanum and determination of manganese by atomic absorption spectrometry in an acetylene–air flame at 279. The test is performed according to PN-Z-04472:2015-10 + Ap1:2015-12; Wavelength / flame: 279.5 nm; acetylene–air oxidising; hollow cathode lamp for Mn. The procedure comprises 6 steps; it is used for: Assessment of the exposure of welders and of workers in metallurgy, in foundries, in the manufacture of alloys and in the processing of manganese ores, Plants using manganese compounds: manufacture of cells, paints, fertilisers, glass, ceramics, wood preservatives, Periodic measurements under the regulation of the Minister of Health — comparison with the NDS of 0.2 mg/m³ (inhalable fraction) and 0.05 mg/m³ (respirable fraction).
At a glance
- Standard: PN-Z-04472:2015-10 + Ap1:2015-12
- Category: Environment
- Procedure steps: 6
- Wavelength / flame: 279.5 nm; acetylene–air oxidising; hollow cathode lamp for Mn
- Calibration curve: 0.30; 0.50; 1.00; 2.00; 4.00; 6.00 µg/ml Mn with about 0.5 % La; correlation coefficient 1.0000
- Limit of quantification: 0.04 µg/ml; limit of determination 0.02 mg/m³ (inhalable) and 0.005 mg/m³ (respirable)
Overview
Manganese is used in metallurgy for the manufacture of ferrous and non-ferrous alloys and of steel; manganese dioxide — in galvanic cells, as an oxidising agent and as an additive to paints; other compounds — in fertilisers, feeds, wood preservatives and in the pharmaceutical, glass and ceramic industries. Occupational exposure is connected with the extraction and the processing of ores, with the manufacture and the casting of alloys and with welding with electrodes containing manganese. Chronic exposure damages the nervous, the respiratory and the reproductive systems; poisoning with manganese clinically resembles Parkinson's disease and is irreversible. The NDS values: 0.2 mg/m³ for the inhalable fraction and 0.05 mg/m³ for the respirable fraction — the assessment of the exposure is carried out separately for the two fractions.
The method of flame atomic absorption spectrometry worked out at CIOP-PIB constitutes a revision of the standard PN-79/Z-04125.02 and has been embodied in the standard PN-Z-04472:2015-10; the limit of determination is 0.02 mg/m³ for the inhalable fraction and 0.005 mg/m³ for the respirable fraction (air volume 600–720 l, final volume of the sample 10 ml). The CIOP-PIB procedure of 2021 determines manganese together with nickel and iron in a single analytical run; its measuring range for manganese of 0.30–8.0 µg/ml corresponds to 0.017–0.44 mg/m³ in the inhalable fraction (dilution 4) and to 0.0042–0.11 mg/m³ in the respirable fraction (without dilution), that is 0.085–2.2 NDS.
Method principle
A known volume of air (up to 720 l) is drawn through cellulose ester membrane filters (0.85 µm): for the inhalable fraction a filter in a holder with a flow rate of not more than 2 l/min, for the respirable fraction, in parallel, a filter in the upper part of a microcyclone with the flow rate proper to the separation of that fraction. Each filter is mineralised in a 50 ml beaker with 2 ml of concentrated HNO₃ on a hot plate (about 140 °C) until the filter has dissolved and the contents have evaporated almost to dryness (several times), 2 ml of 10 % HNO₃ is added, after 30 min it is transferred into a 10 ml flask and made up with 0.1 mol/l HNO₃. The solution of the inhalable fraction is diluted (2.5 ml + 1.5 ml of 10 % HNO₃ + 0.5 ml of lanthanum chloride to 10 ml, k = 4); the solution of the respirable fraction is prepared without dilution with the addition of 0.5 ml of lanthanum chloride. Manganese is determined at 279.5 nm in an oxidising acetylene–air flame, against a calibration curve of 0.30–6.00 µg/ml and with a blank from a clean filter; the concentration X = (c − c0)·V1·k/V [mg/m³], separately for each fraction.
Applications
- Assessment of the exposure of welders and of workers in metallurgy, in foundries, in the manufacture of alloys and in the processing of manganese ores
- Plants using manganese compounds: manufacture of cells, paints, fertilisers, glass, ceramics, wood preservatives
- Periodic measurements under the regulation of the Minister of Health — comparison with the NDS of 0.2 mg/m³ (inhalable fraction) and 0.05 mg/m³ (respirable fraction)
- Personal sampling with parallel sampling of the two fractions of the aerosol
- Determination of Mn together with Ni and Fe from one series of samples (the CIOP-PIB procedure 2021)
Key parameters
| Parameter | Value |
|---|---|
| Wavelength / flame | 279.5 nm; acetylene–air oxidising; hollow cathode lamp for Mn |
| Calibration curve | 0.30; 0.50; 1.00; 2.00; 4.00; 6.00 µg/ml Mn with about 0.5 % La; correlation coefficient 1.0000 |
| Limit of quantification | 0.04 µg/ml; limit of determination 0.02 mg/m³ (inhalable) and 0.005 mg/m³ (respirable) |
| Precision / uncertainty | inhalable fraction 5.42 % / 11.9 %; respirable 5.38 % / 11.8 % |
| Sampling | up to 720 l; inhalable fraction ≤ 2 l/min in a holder; respirable — a microcyclone |
| NDS | inhalable fraction 0.2 mg/m³; respirable fraction 0.05 mg/m³ |
Standard
- Standard number
- PN-Z-04472:2015-10 + Ap1:2015-12
- Title (PL)
- Ochrona czystości powietrza — Oznaczanie manganu i jego związków na stanowiskach pracy metodą płomieniowej absorpcyjnej spektrometrii atomowej
- Title (EN)
- Air purity protection — Determination of manganese and its compounds at workplaces by flame atomic absorption spectrometry
Step-by-step procedure
-
Sampling
In accordance with the principles of PN-Z-04008-7 take two samples in parallel in the breathing zone: the inhalable fraction through a filter in a holder (up to 720 l, flow rate ≤ 2 l/min) and the respirable fraction through a filter in the upper part of a microcyclone (up to 720 l, the flow rate proper to the microcyclone).
-
Mineralisation
Place each filter in a 50 ml beaker, add 2 ml of concentrated HNO₃ and heat it on the hot plate (about 140 °C) until the filter has dissolved and the contents have evaporated almost to dryness; repeat until the sample has completely dissolved. Add 2 ml of 10 % HNO₃, wait about 30 min, transfer it quantitatively with 5–6 portions of 0.1 mol/l HNO₃ into a 10 ml flask and make up to the mark. In parallel mineralise clean filters (the blanks No 1 and No 2).
-
Solutions for the test
Inhalable fraction (No 1): into a 10 ml flask transfer 2.5 ml of the solution after the mineralisation, add 1.5 ml of 10 % HNO₃ and 0.5 ml of lanthanum chloride and make up with 0.1 mol/l HNO₃ (k = 4). Respirable fraction (No 2): into the 10 ml flask with the residue after the mineralisation add 0.5 ml of lanthanum chloride and make up with 0.1 mol/l HNO₃ (without dilution).
-
Calibration curve
On the day of the analysis prepare six working solutions of 0.30–6.00 µg/ml Mn (0.5 ml of lanthanum chloride and 2 ml of 10 % HNO₃ per 10 ml) and a zero sample. Make three measurements of the absorbance at 279.5 nm in an oxidising acetylene–air flame, zeroing with 0.1 mol/l HNO₃; automatic calibration of the instrument is permitted.
-
Measurement
Measure the absorbance of the solutions No 1 and No 2 and of the blanks under the conditions as for the calibration and read the concentrations from the curve. Where the range is exceeded repeat the determination with a smaller volume of the solution, taking the dilution factor into account.
-
Calculation and assessment
For each fraction calculate X = (c − c0)·V1·k/V [mg/m³] (V1 = 10 ml; k = 4 for the inhalable fraction, 1 for the respirable one; V — the volume of air in l). Compare it separately with the NDS of 0.2 mg/m³ (inhalable) and 0.05 mg/m³ (respirable).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Atomic absorption spectrometer with an acetylene–air burner | with a hollow cathode lamp for manganese (in the validation a Thermo SOLAAR M) | — |
| Suction pump (personal sampler) | a constant volume flow rate; ≤ 2 l/min for the inhalable fraction, according to the manufacturer of the microcyclone for the respirable one | — |
| Filter holder (inhalable fraction) and a microcyclone of plastics (respirable fraction) | commercially available samplers of the fractions of the aerosol | — |
| Cellulose ester membrane filters | pore diameter 0.85 µm | — |
| Hot plate | mineralisation in 50 ml beakers at about 140 °C under a fume cupboard | — |
| Volumetric flasks of 10 ml and 100 ml, pipettes, polyethylene vessels for the standards | borosilicate glass washed with detergent, 1 mol/l HNO₃ and deionised water | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Concentrated nitric acid 65 % (ρ = 1.39 g/ml), analytical grade | — | mineralisation of the filters |
| Nitric acid 10 % (v/v) — solution I | — | dissolution of the residue, addition to the standards and to the solution of the inhalable fraction |
| Nitric acid 0.1 mol/l — solution II | — | making up of the flasks, intermediate solution, zeroing of the spectrometer |
| Lanthanum chloride solution (26.8 g of LaCl₃·7H₂O in 73.2 ml of 0.1 mol/l HNO₃; about 10 % La) | — | spectral buffer (0.5 ml per 10 ml of solution) |
| Basic standard solution of manganese for AAS 1 mg/ml | — | intermediate solution 10 µg/ml (stable for 14 days) and working solutions 0.30–6.00 µg/ml |
| Dissolved acetylene of class A (PN-C-84905) | — | combustible gas for the flame |
| Doubly distilled or deionised water | — | preparation of the solutions |
Health and safety (OHS)
- Operations with concentrated nitric acid — in protective clothing and under an efficient fume cupboard; the hot mineralisation evolves nitrogen oxides
- Operate the acetylene and the AAS burner according to the instructions of the manufacturer; check the leak-tightness of the gas installation
- Collect the residues of the solutions of the reagents and of the standards in containers and hand them over for disposal to authorised institutions
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-Z-04472:2015-10 + Ap1:2015-12 — “Air purity protection — Determination of manganese and its compounds at workplaces by flame atomic absorption spectrometry”.
How does this method work?
A known volume of air (up to 720 l) is drawn through cellulose ester membrane filters (0.85 µm): for the inhalable fraction a filter in a holder with a flow rate of not more than 2 l/min, for the respirable fraction, in parallel, a filter in the upper part of a microcyclone with the flow rate proper to the separation of that fraction.
What is the measuring range and accuracy?
Wavelength / flame: 279.5 nm; acetylene–air oxidising; hollow cathode lamp for Mn; Calibration curve: 0.30; 0.50; 1.00; 2.00; 4.00; 6.00 µg/ml Mn with about 0.5 % La; correlation coefficient 1.0000; Limit of quantification: 0.04 µg/ml; limit of determination 0.02 mg/m³ (inhalable) and 0.005 mg/m³ (respirable); Precision / uncertainty: inhalable fraction 5.42 % / 11.9 %; respirable 5.38 % / 11.8 %.
How long does the test take?
The procedure comprises 6 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Atomic absorption spectrometer with an acetylene–air burner, Suction pump (personal sampler), Filter holder (inhalable fraction) and a microcyclone of plastics (respirable fraction), Cellulose ester membrane filters, Hot plate, Volumetric flasks of 10 ml and 100 ml, pipettes, polyethylene vessels for the standards.
Where is this test used?
Assessment of the exposure of welders and of workers in metallurgy, in foundries, in the manufacture of alloys and in the processing of manganese ores; Plants using manganese compounds: manufacture of cells, paints, fertilisers, glass, ceramics, wood preservatives; Periodic measurements under the regulation of the Minister of Health — comparison with the NDS of 0.2 mg/m³ (inhalable fraction) and 0.05 mg/m³ (respirable fraction); Personal sampling with parallel sampling of the two fractions of the aerosol; Determination of Mn together with Ni and Fe from one series of samples (the CIOP-PIB procedure 2021).
What safety precautions apply?
Operations with concentrated nitric acid — in protective clothing and under an efficient fume cupboard; the hot mineralisation evolves nitrogen oxides; Operate the acetylene and the AAS burner according to the instructions of the manufacturer; check the leak-tightness of the gas installation; Collect the residues of the solutions of the reagents and of the standards in containers and hand them over for disposal to authorised institutions.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-Z-04472.