🌡️ Total mercury in waste gases — manual method according to PN-EN 13211

Environment PN-EN 13211

In short

Manual reference method for the determination of total mercury in waste gases: isokinetic sampling through a heated probe and filter, absorption of gaseous mercury in impingers, determination by the cold vapour technique. The test is performed according to PN-EN 13211:2006 (wersja polska; wprowadza EN 13211:2001 [IDT] i EN 13211:2001/AC:2005 [IDT]); Validated range: 0.001–0.5 mg/m³ (1–500 µg/m³) — according to secondary sources, not from the text of the standard. The procedure comprises 7 steps; it is used for: Measurements of mercury emissions from waste incineration and co-incineration plants as the reference method, Checking compliance with mercury emission limit values in integrated permits, Calibration and checking of automated measuring systems for mercury (AMS) with the reference method.

At a glance

  • Standard: PN-EN 13211:2006 (wersja polska; wprowadza EN 13211:2001 [IDT] i EN 13211:2001/AC:2005 [IDT])
  • Category: Environment
  • Procedure steps: 7
  • Quantity determined: mass concentration of TOTAL mercury — the sum of mercury in the gas phase and mercury bound to dust
  • Validated range: 0.001–0.5 mg/m³ (1–500 µg/m³) — according to secondary sources, not from the text of the standard
  • Absorption solution I — one of the TWO permitted, to be chosen: permanganate: 2 % KMnO₄ in 10 % H₂SO₄

Overview

WHAT THIS STANDARD IS. PN-EN 13211:2006 is a REFERENCE method (SRM) for the determination of the mass concentration of total mercury in gases discharged through ducts and stacks. "Total mercury" means here the sum of mercury in the gas phase and mercury bound to dust — both fractions go into one result, although they are collected separately: dust on the filter, gaseous mercury in the impingers. PKN card: the standard is CURRENT, Polish version of 2006, it implements EN 13211:2001; the earlier edition PN-EN 13211:2002 (English version) was replaced by it. In accreditation scopes it is cited by 25 laboratories in 63 entries.

SCOPE OF APPLICATION. The method is validated for total mercury concentrations from 0.001 to 0.5 mg/m³, i.e. from 1 to 500 µg/m³. The lower limit has practical significance: the emission limit values for waste incineration plants lie in the same order of magnitude, so the method works close to its lower end and the quality of sampling decides the result more than the analysis itself.

TWO ABSORPTION SOLUTIONS — A CHOICE THAT MUST BE RECORDED. The standard permits two solutions: permanganate (2 % KMnO₄ in 10 % H₂SO₄) or dichromate (4 % K₂Cr₂O₇ in 20 % HNO₃). Both oxidise mercury and bind it in the solution, but they differ in how the sample is prepared for analysis and in their susceptibility to interferences. This is the point at which two laboratories carrying out "the same test according to PN-EN 13211" may work differently — and that is why it is worth indicating in the report and in the accreditation scope which solution was used. The standard number alone does not settle this.

CONDITIONS THAT DECIDE THE VALIDITY OF THE MEASUREMENT. Sampling must be ISOKINETIC — the suction velocity in the nozzle equal to the gas velocity in the duct — and the permissible range of isokineticity given for this method is 95–115 %. Outside this range the sample is not representative of the dust, and therefore not of the mercury bound to the dust either. The second criterion is an ABSORPTION EFFICIENCY above 95 %, checked by comparing the mercury content in successive impingers (the procedure described in the secondary sources we used; we did not have the text of the standard): if the last impinger contains significantly more than a trace, this means breakthrough and the result is underestimated. Both numbers are conditions of validity, not parameters of the result — a measurement that meets them gives a result, a measurement that does not meet them gives none.

FINAL DETERMINATION. The mercury content of the solutions from the impingers and of the rinsings, as well as the mercury from the filter, is determined by the COLD VAPOUR technique — atomic absorption spectrometry (CVAAS) or atomic fluorescence spectrometry (CVAFS). In the process, mercury is reduced to the atomic form and carried into the measuring cell in a stream of carrier gas; hence the name "cold vapour" — atomisation requires neither a flame nor a furnace.

TOTAL MERCURY VERSUS SPECIATION. The standard gives the sum and does not separate the forms of mercury. In measurement practice, elemental and bound (oxidised) mercury are distinguished, and their proportions depend on the fuel and on the flue gas cleaning system — but both enter the result according to PN-EN 13211. A laboratory from which separation of the forms is expected needs a different solution than this standard alone.

WHAT WE DO NOT GIVE. We did not have the text of PN-EN 13211:2006 or EN 13211:2001 — the standard is sold for a fee. The description comes from SECONDARY sources: standard cards of standardisation bodies, a report on emission measurements carried out with this method, and a publication on an interlaboratory comparison conducted according to EN 13211 and EN 14385. We therefore do not give: the volume and number of impingers, the filtration temperature, the minimum volume of the gas sample, the sampling time or the way of calculating the limit of quantification for a particular set-up. The concentrations of the absorption solutions and the criteria (isokineticity, absorption efficiency) come from the same secondary sources, not from the text of the standard — and should be treated as such.

Method principle

The gas sample is taken isokinetically through a heated probe and a filter that retains dust. Mercury in the gas phase is absorbed in a series of impingers containing an oxidising solution — permanganate or dichromate — which converts mercury into a form bound in the solution. Mercury retained on the filter is brought into solution separately. The mercury content of all fractions is determined by the cold vapour technique (CVAAS or CVAFS), and the result is given as the sum: the concentration of total mercury in the gas, related to the sample volume under standard conditions.

Applications

Key parameters

ParameterValue
Quantity determinedmass concentration of TOTAL mercury — the sum of mercury in the gas phase and mercury bound to dust
Validated range0.001–0.5 mg/m³ (1–500 µg/m³) — according to secondary sources, not from the text of the standard
Absorption solution I — one of the TWO permitted, to be chosenpermanganate: 2 % KMnO₄ in 10 % H₂SO₄
Absorption solution IIdichromate: 4 % K₂Cr₂O₇ in 20 % HNO₃ — an alternative to I; the choice affects the further handling of the sample and must be recorded
Practical difference between the solutionsthe permanganate solution must be prepared in a permanent laboratory (addition of concentrated hydrochloric acid); with dichromate, the nitric acid solution is prepared in the laboratory and the dichromate can be added at the measurement site — secondary source (MID), not the text of the standard
PROHIBITION OF MIXING SOLUTIONSthe permanganate solution MUST NOT be combined with the absorption solutions for other gaseous metals (determined according to PN-EN 14385): contact of nitric acid with sulphuric acid causes gas evolution, a rise in pressure and risks bursting the vessel
When to use which solution — conditions of choiceWE HAVE NOT ESTABLISHED THIS. The secondary sources describe the differences in preparation and the prohibition of mixing, but do not state whether the gas composition (reducing substances, halogens) rules out either solution. We did not have the text of the standard (state as of 22-09-2026)
Validity condition — isokineticity95–115 %; outside this range the sample is not representative of the dust, and therefore not of the mercury bound to it either
Validity condition — absorption efficiency> 95 %; checked by comparing the mercury content in successive impingers (breakthrough underestimates the result)
Final determinationcold vapour technique — CVAAS or CVAFS; atomisation without a flame and without a furnace
What the standard does NOT giveseparation of the forms of mercury (elemental versus oxidised) — the result is a sum; speciation requires a different solution
Status of the standardCURRENT — PN-EN 13211:2006 (Polish version), implements EN 13211:2001; it replaced PN-EN 13211:2002 (English version)
Scale of citations36 laboratories, 102 entries in accreditation scopes — across all forms of notation (10 different; the form PN-EN 13211+AC:2006 alone gives 63 entries) (measurement on the database of scopes, 22-09-2026). The number was counted in the reference document field of the scopes database — the same field in which the "Laboratories" tab searches

Standard

Standard number
PN-EN 13211:2006 (wersja polska; wprowadza EN 13211:2001 [IDT] i EN 13211:2001/AC:2005 [IDT])
Title (PL)
Jakość powietrza — Emisja ze źródeł stacjonarnych — Manualna metoda oznaczania stężenia rtęci ogólnej
Title (EN)
Air quality — Stationary source emissions — Manual method of determination of the concentration of total mercury

Step-by-step procedure

  1. Survey of the conditions at the emission source

    Establish the velocity and temperature of the gas, the dust content and the moisture, and select the nozzle for the sampling conditions. Check whether the expected concentration is within the validated range of the method.

  2. Choosing the absorption solution and recording it

    Decide whether the permanganate or the dichromate solution will be used. Record the choice in the measurement sheet — two laboratories working according to the same standard may use different solutions, and the standard number alone does not settle this.

  3. Isokinetic sampling

    Take the sample keeping the isokineticity within the permissible range (95–115 %). A deviation outside this range invalidates the measurement as regards the mercury bound to dust.

    ⏱ Time: depending on the concentration and the required sample volume

  4. Securing the dust fraction

    Bring the filter with the retained dust into solution by a separate route — mercury bound to dust enters the result together with gaseous mercury, but is not collected together with it.

  5. Checking the absorption efficiency

    Compare the mercury content in successive impingers. The absorption efficiency should exceed 95 %; a significant content in the last impinger means breakthrough and an underestimated result.

  6. Determination by the cold vapour technique

    Reduce mercury to the atomic form and determine it in the stream of carrier gas by CVAAS or CVAFS — separately in the solutions from the impingers, in the rinsings and in the solution from the filter.

  7. Calculation and report

    Add up the fractions, relate them to the sample volume under standard conditions and give the result as the concentration of total mercury. In the report, indicate the absorption solution used, the isokineticity achieved and the check of the absorption efficiency — without them the number cannot be verified.

Required equipment and apparatus

EquipmentExampleIndicative price
Isokinetic sampling setheated probe, nozzle selected for the gas velocity, system for measuring velocity and pressure; isokineticity kept within the range 95–115 %—
Filter with heated holderretains the dust together with the mercury bound to it; this fraction is brought into solution separately and added to the result—
Series of impingersfilled with permanganate or dichromate solution; the last impinger is used to check whether breakthrough has occurred—
Spectrometer for the cold vapour techniqueCVAAS or CVAFS — determination of mercury in the solutions from the impingers, in the rinsings and in the solution from the filter—

Reagents, media and consumables

ReagentCASDetails
Potassium permanganate7722-64-7Absorption solution I — 2 % KMnO₄ in 10 % H₂SO₄; oxidises mercury and binds it in the solution
Sulphuric acid7664-93-9Medium of absorption solution I (concentration 10 %)
Potassium dichromate7778-50-9Absorption solution II — 4 % K₂Cr₂O₇ in 20 % HNO₃; an alternative to the permanganate solution
Nitric acid7697-37-2Medium of absorption solution II (concentration 20 %)
Reducing agent for the cold vapour technique—Tin(II) chloride or sodium borohydride — converts mercury into the atomic form before measurement

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN 13211:2006 (wersja polska; wprowadza EN 13211:2001 [IDT] i EN 13211:2001/AC:2005 [IDT]) — “Air quality — Stationary source emissions — Manual method of determination of the concentration of total mercury”.

How does this method work?

The gas sample is taken isokinetically through a heated probe and a filter that retains dust. Mercury in the gas phase is absorbed in a series of impingers containing an oxidising solution — permanganate or dichromate — which converts mercury into a form bound in the solution.

What is the measuring range and accuracy?

Quantity determined: mass concentration of TOTAL mercury — the sum of mercury in the gas phase and mercury bound to dust; Validated range: 0.001–0.5 mg/m³ (1–500 µg/m³) — according to secondary sources, not from the text of the standard; Absorption solution I — one of the TWO permitted, to be chosen: permanganate: 2 % KMnO₄ in 10 % H₂SO₄; Absorption solution II: dichromate: 4 % K₂Cr₂O₇ in 20 % HNO₃ — an alternative to I; the choice affects the further handling of the sample and must be recorded.

How long does the test take?

The procedure comprises 7 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Isokinetic sampling set, Filter with heated holder, Series of impingers, Spectrometer for the cold vapour technique.

Where is this test used?

Measurements of mercury emissions from waste incineration and co-incineration plants as the reference method; Checking compliance with mercury emission limit values in integrated permits; Calibration and checking of automated measuring systems for mercury (AMS) with the reference method; Participation in interlaboratory comparisons conducted according to EN 13211 and EN 14385; Indicating in the report which of the two permitted absorption solutions was used.

What safety precautions apply?

PROHIBITION OF MIXING: the permanganate solution in sulphuric acid must not be combined with the absorption solutions for other gaseous metals (nitric acid) — contact of the two acids causes gas evolution and a rise in pressure that risks bursting the vessel. This applies to measurements in which mercury is determined in parallel with metals according to PN-EN 14385, i.e. a typical situation; Mercury and its vapours are toxic — the solutions after the measurement contain mercury in bound form and are hazardous waste, handed over for disposal; The permanganate solution in sulphuric acid is strongly oxidising and corrosive; preparation requires adding the acid to water, never the other way round; Potassium dichromate is a carcinogenic and mutagenic substance, and its solutions in nitric acid are corrosive — work only in a fume cupboard, in chemically resistant gloves; Sampling at the emission source takes place on the measurement platform with hot gases — the rules for working at height and protection against burns from the probe apply; A result without the stated isokineticity and without the check of the absorption efficiency cannot be verified — both quantities are conditions of validity of the measurement, not an addition to the report.

Which laboratory can perform this test?

The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 13211.