⚗️ Dissolved sulfide in water and wastewater — filtration through 0.45 µm, preservation with ascorbate, stripping with nitrogen at pH 4 and determination with methylene blue at 665 nm
In short
A photometric method for determining dissolved sulfide in the range from 0. The test is performed according to ISO 10530:1992; Measuring range: from 0.04 mg/l to 1.5 mg/l of sulfide; higher concentrations — after reducing the volume of the sample and diluting. The procedure comprises 12 steps (about 1 days 57 min–1 days 1 h 7 min in total); it is used for: Control of municipal wastewater and sewers for the release of hydrogen sulfide and the sulfate corrosion of concrete collectors, Industrial wastewater from tanneries, paper mills (the sulfate process), coking plants, refineries and rubber-processing works, Natural waters under anaerobic conditions — groundwater, bottom water of reservoirs, landfill leachate.
At a glance
- Standard: ISO 10530:1992
- Category: Physicochemistry
- Procedure steps: 12
- Total time of stages: 1 days 57 min–1 days 1 h 7 min
- Quantity determined: dissolved sulfide — that is, what passes through a 0.45 µm filter and is stripped with nitrogen at pH 4; not the total sulfur content and not the sulfide bound to the suspended matter
- Measuring range: from 0.04 mg/l to 1.5 mg/l of sulfide; higher concentrations — after reducing the volume of the sample and diluting
- Type of samples: wastewater and natural waters which can be filtered. Waters which cannot be filtered must not be analysed by this method — for them one determines the sulfide readily released at pH 4 (ISO 13358)
Overview
Sulfides in water and wastewater are troublesome for three reasons at once: they are toxic to aquatic organisms, on acidification they release hydrogen sulfide — a gas deadly to sewer staff — and they are responsible for the sulfate corrosion of concrete sewers. The analytical trouble is different: sulfide is unstable. The standard itself states that because of the instability of dissolved sulfide and because the filtration of an unpreserved sample is an integral part of the method, it was not possible to carry out classical interlaboratory trials with samples sent out — instead a comparative trial was performed in a single laboratory, by operators from different institutes, on a solution with a sulfide concentration of 1 mg/l. That one sentence from the clause on precision says more about this method than the whole procedure: the sample cannot be put off until later.
Scope and limits of application. The method is intended for determining dissolved sulfide in the range of mass concentrations from 0.04 mg/l to 1.5 mg/l; higher concentrations are determined by reducing the volume of the sample used and by dilution. It applies to wastewater and natural waters which can be filtered. This limitation is hard: waters which cannot be filtered in the manner described in clause 6 of the standard must not be analysed by this method — for them one determines the sulfide readily released at pH 4, that is what ISO 13358:1997 describes (Dosage des sulfures aisément libérables). With the procedure as laid down, the determination of the part of the sulfide coming from polysulfides is incomplete. There is no interference from: cyanides up to 2 mg/l, iodides up to 20 mg/l, thiosulfates up to 900 mg/l, thiocyanates up to 900 mg/l and sulfites up to 700 mg/l; nor from carbon disulfide below 10 mg/l and ethanethiol (formerly: ethyl mercaptan) below 1 mg/l.
Which standard is the right one for a Polish laboratory. The national candidate, the series PN-C-04566 „Woda i ścieki — Badania zawartości siarki i jej związków" [Water and wastewater — Tests for the content of sulfur and its compounds], is not suitable as a reference document: in the PKN catalogue all its sheets have the status "Withdrawn". This also applies to all the sheets devoted to sulfide itself — PN-C-04566-02:1982 (colorimetric method with thiofluorescein and o-hydroxymercuribenzoic acid), PN-C-04566-03:1982 (thiomercurimetric method), PN-C-04566-04:1974 (undissociated hydrogen sulfide), PN-C-04566-05:1981 (ion-selective electrode) and PN-C-04566-13:1985 (mercurimetric method with preliminary concentration) — for none of them does the catalogue name a superseding standard. For example the card of sheet -03:1982 (titration with the salt of o-hydroxymercuribenzoic acid against dithizone, the sum of hydrogen sulfide and soluble sulfides above 0.04 mg/l) gives the date of publication 20 April 1982 and the date of withdrawal 16 January 2006, and there is no "Superseded by" field at all — the sheet disappeared from the PN collection without a successor. The only sheet of that series which did obtain a successor is PN-C-04566-09:1974 (sulfates by the gravimetric method), superseded by PN-ISO 9280:2002. At the same time ISO 10530 has never been adopted into the collection of Polish Standards — a search of the PKN catalogue for "ISO 10530" returns no item at all. The practical conclusion: the reference document is ISO 10530:1992 itself, and the entry in the accreditation scope must reflect that (an ISO standard, not a PN), or else the laboratory describes the same course of action in its own test procedure with a reference to ISO 10530.
Status of ISO 10530:1992. The standard is current. First edition of 15 September 1992, prepared by ISO/TC 147 "Water quality", subcommittee SC 2 "Physical, chemical and biochemical methods"; Annex A (iodometric titration for standardising sulfide solutions) is an integral part of the standard. The DIN Media catalogue marks it with the tag [CURRENT] as ISO 10530:1992-09, 9 pages, original in English. It is neither withdrawn nor superseded.
What this description does not contain. We worked on a preview of the standard in the French version — it covers practically the whole content: scope, interferences, principle, reagents, apparatus, sampling and preparation of samples, procedure, calibration curve, calculation, part of the clause on precision and the test report. We did not, however, see Table 1 with the numerical precision characteristics or the full text of Annex A, and for that reason we give no figures on the repeatability or the reproducibility of this method. Nor do we give an official Polish title, because none exists — the Polish title in the heading is a working translation.
THE POLISH TITLE IN THIS ENTRY IS OUR WORKING TRANSLATION, not an official title: the standard has not been adopted into the collection of Polish Standards.
Method principle
The water sample is filtered in order to separate the suspended matter and the sparingly soluble sulfides, and the sulfide in the filtrate is preserved by adding an ascorbate solution. The filtrate is transferred into a reaction flask with a phthalate buffer of pH 4.0 and the sulfide is stripped from it with a stream of nitrogen, being carried into a scrubber with an aqueous solution of zinc acetate, where it is bound as zinc sulfide. In the scrubber, after the addition of a solution of N,N-dimethyl-1,4-phenylenediammonium chloride in sulfuric acid, leucomethylene blue is formed, which by the addition of iron(III) ions is oxidised to methylene blue. The absorbance of the dye formed is measured at a wavelength of 665 nm against water, and the concentration is read from the linear section of the calibration curve.
Applications
- Control of municipal wastewater and sewers for the release of hydrogen sulfide and the sulfate corrosion of concrete collectors
- Industrial wastewater from tanneries, paper mills (the sulfate process), coking plants, refineries and rubber-processing works
- Natural waters under anaerobic conditions — groundwater, bottom water of reservoirs, landfill leachate
- Assessment of the efficiency of aeration and of nitrate dosing in a sewer network as a means against the formation of sulfide
- Testing of sulfide (medicinal) waters and control of installations in which iron sulfide is deposited
- Diagnosis of the causes of blackening of sludge and of the precipitation of metal sulfides in treatment plants
Key parameters
| Parameter | Value |
|---|---|
| Quantity determined | dissolved sulfide — that is, what passes through a 0.45 µm filter and is stripped with nitrogen at pH 4; not the total sulfur content and not the sulfide bound to the suspended matter |
| Measuring range | from 0.04 mg/l to 1.5 mg/l of sulfide; higher concentrations — after reducing the volume of the sample and diluting |
| Type of samples | wastewater and natural waters which can be filtered. Waters which cannot be filtered must not be analysed by this method — for them one determines the sulfide readily released at pH 4 (ISO 13358) |
| Wavelength | 665 nm; measurement against water, cells of 1 cm optical path |
| Stripping | nitrogen of 99.996 % purity, flow rate 40 l/h: 2 min before the filtrate is introduced and 30 min with the filtrate in the reaction flask |
| Non-interfering substances (up to the concentrations given) | cyanides 2 mg/l, iodides 20 mg/l, thiosulfates 900 mg/l, thiocyanates 900 mg/l, sulfites 700 mg/l; carbon disulfide < 10 mg/l, ethanethiol < 1 mg/l. Sulfide originating from polysulfides is determined incompletely |
| Time limit for performance | preserved filtrates are analysed as soon as possible, at the latest 24 h after the sample was taken; the time of filtration under nitrogen pressure must not exceed 5 min |
| Result | rounded to 0.01 mg and given with at most two significant figures (the example from the standard: "Dissolved sulfide: 0.55 mg/l") |
Standard
- Standard number
- ISO 10530:1992
- Title (PL)
- Jakość wody — Oznaczanie siarczków rozpuszczonych — Metoda fotometryczna z błękitem metylenowym
- Title (EN)
- Water quality — Determination of dissolved sulfide — Photometric method using methylene blue
Step-by-step procedure
-
Deciding whether the method fits at all
Check whether the water can be filtered in the manner described in clause 6 of the standard. If not — do not analyse it by this method; for such waters one determines the sulfide readily released at pH 4 according to ISO 13358. If it is not known whether the concentration will fall within the range 0.04–1.5 mg/l, the standard recommends filtering and preserving several portions of the sample so that smaller amounts can also be determined.
-
Filtration and preservation — waters easy to filter
Into a 50 ml volumetric flask pipette 5 ml of the ascorbate buffer. Draw the water into a 50 ml three-ring piston syringe, fit a single-use 0.45 µm filter and filter the water into the volumetric flask up to the mark.
-
Filtration and preservation — waters difficult to filter
Into a 50 ml volumetric flask introduce 5 ml of the ascorbate buffer. Flush the flask and the pressure filtration set with nitrogen for about 10 min. Fill the pressure vessel completely with the water under test and filter it into the volumetric flask up to the mark at a nitrogen pressure of up to 2 bar. Make the connection between the set and the flask so as to limit as far as possible the ingress of air. The filtration time must not exceed 5 min.
-
Time limit for the analysis
Analyse the preserved filtrates as soon as possible, at the latest 24 h after the sample was taken. This is not an organisational recommendation but a condition of the validity of the result — dissolved sulfide is unstable.
-
Preparing the stripping apparatus
Pour 25 ml of the phthalate buffer pH 4.0 into the reaction flask. Pour 20 ml of the zinc acetate solution into the absorption tube (scrubber). Assemble the apparatus and pass nitrogen through the solution at a rate of 40 l/h for 2 min in order to remove the air from the system.
-
Introducing the filtrate and stripping the sulfide
Transfer the filtrate into the reaction flask through the dropping funnel, rinse it with a small amount of water, and then pass nitrogen through the solution at a rate of 40 l/h for 30 min. At pH 4 the sulfide is converted into hydrogen sulfide, carried along by the nitrogen and retained in the scrubber as zinc sulfide.
-
Developing the colour
Take the absorption tube off the apparatus and through the side joint add 10 ml of the reagent for forming the coloured complex, and then 1 ml of the ammonium iron(III) sulfate solution. Fill the absorption tube completely with water, stopper it, shake and wait 10 min.
-
Measuring the absorbance
Pour the solution into a 100 ml volumetric flask, rinse the absorption tube carefully with a small amount of water and add the washings to the solution in the flask. Make up with water to the mark and measure the absorbance at 665 nm against water. If the mass concentration exceeds 1.5 mg/l of sulfide, repeat the determination using a smaller portion of the preserved filtrate.
-
Blank
Carry out a blank by exactly the same course of action. The response of the blank must not differ appreciably from the calculated value A₍s0₎ — the intercept of the calibration curve.
-
Calibration curve
Into seven 100 ml volumetric flasks introduce 20 ml of the zinc acetate solution each. Into six of them pipette respectively 4, 6, 8, 10, 12 and 14 ml of the sulfide standard solution; the seventh serves as the blank. Into each flask add 10 ml of the colour reagent and 1 ml of the ammonium iron(III) sulfate solution, dilute with water to about 40 ml, stopper, shake and make up with water to the mark. After 10–20 min measure the absorbance at 665 nm against water in the reference cell. The mass concentrations of sulfide in these solutions are approximately 0 (blank), 0.2, 0.3, 0.4, 0.5, 0.6 and 0.7 mg/l, and their exact values are determined iodometrically according to Annex A. The curve is not always linear over the whole range — only its linear section may be used for the calculations. Every spectrometer and every optical path requires its own curve.
-
Calculating and rounding the result
The concentration of dissolved sulfide is calculated as (A₍s₎ − A₍s0₎)·f / (h·V), where A₍s₎ is the absorbance of the sample, A₍s0₎ the calculated absorbance of the blank, h the slope of the calibration curve (sensitivity in l/mg), f = 100 ml (conversion factor), V = 45 ml (the volume of filtrate used for the analysis, that is the 50 ml of the flask less the 5 ml of ascorbate buffer). Every dilution must be taken into account in the calculation. Round the result to 0.01 mg and give it with at most two significant figures.
-
Supervision of the calibration
Check the slope of the curve and its intercept with the ordinate axis at regular intervals in order to detect appreciable deviations; such checks are particularly advisable after starting a new batch of reagent.
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Set for filtering the sample | a three-ring piston syringe of 50 ml capacity with a single-use filter of 0.45 µm porosity; for waters difficult to filter — a set for pressure filtration with a 0.45 µm membrane, working under nitrogen up to 2 bar | — |
| Apparatus for gas stripping | a 250 ml reaction flask with a side ground joint for connecting a 100 ml dropping funnel, a gas inlet tube reaching the bottom of the flask and ending in a ball Ø 16 mm with holes Ø 0.5 mm, a condenser set vertically or a rising column, and an absorption tube (scrubber) | — |
| Nitrogen supply with a flow meter | nitrogen of 99.996 % (m/m) purity and a flow meter suitable for a rate of 40 l/h | — |
| Spectrometer or filter photometer | permitting the measurement of absorbance at 665 nm; cells of 1 cm optical path | — |
| pH meter with a suitable electrode | for checking the pH of the phthalate (4.0 ± 0.1) and ascorbate (10 ± 0.1) buffers and for the electrometric indication with coloured samples | — |
| Volumetric glassware | volumetric flasks of 50, 100, 500 and 1000 ml, a 25 ml measuring cylinder, graduated pipettes of 1 and 10 ml, one-mark pipettes of 1, 2, 5, 10, 20, 50 and 100 ml, dispensers and microsyringes | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Oxygen-free water | — | distilled water or water of equivalent purity from which the oxygen has been removed by a suitable means — by boiling or by sparging with nitrogen. Without this step the sulfide will be oxidised in the course of the determination |
| Sulfuric acid | 7664-93-9 | ρ(H₂SO₄) = 1.84 g/ml; a component of the colour reagent and of the iron(III) salt solution |
| Sodium hydroxide — solution 32 % (m/m) | 1310-73-2 | c(NaOH) ≈ 10 mol/l; for bringing the ascorbate buffer to pH 10 |
| Zinc acetate — solution | 5970-45-6 | dissolve 20 g of zinc acetate dihydrate [Zn(CH₃COO)₂·2H₂O] in water and make up to 1 l. A turbidity may appear — it does not interfere with the determination. The solution fills the scrubber and binds the stripped hydrogen sulfide as zinc sulfide |
| Phthalate buffer pH 4.0 ± 0.1 | 877-24-7 | 80 g of potassium hydrogen phthalate (C₈H₅KO₄) in 920 ml of water; check the pH and if necessary bring it to 4.0 with dilute NaOH or HCl [e.g. c = 1 mol/l]. It is this that sets the pH at which the sulfide is stripped |
| Ascorbate buffer pH 10 ± 0.1 | 50-81-7 | 10 g of ascorbic acid in 90 ml of water, brought with sodium hydroxide solution to pH 10; close the bottle immediately. Prepare directly before use — it serves to preserve the sulfide in the filtrate |
| Reagent for forming the coloured complex | 536-46-9 | suspend 2 g of N,N-dimethyl-1,4-phenylenediammonium chloride (C₈H₁₄Cl₂N₂) in 200 ml of water in a 1000 ml flask, carefully add 200 ml of sulfuric acid, cool and make up with water to the mark |
| Ammonium iron(III) sulfate — solution | 7783-83-7 | 50 g of ammonium iron(III) sulfate dodecahydrate [NH₄Fe(SO₄)₂·12H₂O] in a 500 ml flask, add 10 ml of sulfuric acid and carefully make up with water. It oxidises leucomethylene blue to methylene blue |
| Sodium sulfide — stock solution | 1313-84-4 | an amount of hydrated sodium sulfide [Na₂S·xH₂O, x = 7–9] corresponding to about 0.5 g of sulfur in the form of sulfide, with a thiosulfate content < 0.5 %, dissolved in water to 1000 ml. Stable for 2–3 days; the exact concentration is determined before use by the iodometric method according to Annex A of the standard |
| Sodium sulfide — standard solution | — | 10 ml of the stock solution made up with water to 1000 ml; 1 ml contains about 5 µg of sulfide. Prepare directly before use, the exact concentration — iodometrically |
Health and safety (OHS)
- The acidified sample and the reaction flask itself give off hydrogen sulfide — a strongly toxic gas which at higher concentrations paralyses the sense of smell, so that the warning smell of rotten eggs ceases to protect. Carry out the whole stripping stage in an efficient fume cupboard.
- The nitrogen outlet from the scrubber must end in the fume cupboard; over the 30 min of stripping the stream of 40 l/h carries the whole stripped amount of hydrogen sulfide.
- The colour reagent contains 200 ml of concentrated sulfuric acid per litre — prepare it by carefully adding the acid to the suspension in water and cool it before making up to the mark.
- N,N-dimethyl-1,4-phenylenediammonium chloride is sensitising and irritant — weigh it wearing gloves, in a fume cupboard, avoiding dust.
- Sodium sulfide is corrosive and in contact with acids releases hydrogen sulfide — do not acidify the stock solutions outside the apparatus and do not pour them into the drain.
- Pressure filtration is carried out under nitrogen up to 2 bar — use vessels intended for work under pressure and a shield; nitrogen in a closed room displaces oxygen.
Frequently asked questions
Which standard describes this test?
The test is performed according to ISO 10530:1992 — “Water quality — Determination of dissolved sulfide — Photometric method using methylene blue”.
How does this method work?
The water sample is filtered in order to separate the suspended matter and the sparingly soluble sulfides, and the sulfide in the filtrate is preserved by adding an ascorbate solution. The filtrate is transferred into a reaction flask with a phthalate buffer of pH 4.
What is the measuring range and accuracy?
Quantity determined: dissolved sulfide — that is, what passes through a 0.45 µm filter and is stripped with nitrogen at pH 4; not the total sulfur content and not the sulfide bound to the suspended matter; Measuring range: from 0.04 mg/l to 1.5 mg/l of sulfide; higher concentrations — after reducing the volume of the sample and diluting; Type of samples: wastewater and natural waters which can be filtered. Waters which cannot be filtered must not be analysed by this method — for them one determines the sulfide readily released at pH 4 (ISO 13358); Wavelength: 665 nm; measurement against water, cells of 1 cm optical path.
How long does the test take?
The times given for the individual stages add up to approximately 1 days 57 min–1 days 1 h 7 min. The procedure comprises 12 steps.
What equipment is required?
Set for filtering the sample, Apparatus for gas stripping, Nitrogen supply with a flow meter, Spectrometer or filter photometer, pH meter with a suitable electrode, Volumetric glassware.
Where is this test used?
Control of municipal wastewater and sewers for the release of hydrogen sulfide and the sulfate corrosion of concrete collectors; Industrial wastewater from tanneries, paper mills (the sulfate process), coking plants, refineries and rubber-processing works; Natural waters under anaerobic conditions — groundwater, bottom water of reservoirs, landfill leachate; Assessment of the efficiency of aeration and of nitrate dosing in a sewer network as a means against the formation of sulfide; Testing of sulfide (medicinal) waters and control of installations in which iron sulfide is deposited; Diagnosis of the causes of blackening of sludge and of the precipitation of metal sulfides in treatment plants.
What safety precautions apply?
The acidified sample and the reaction flask itself give off hydrogen sulfide — a strongly toxic gas which at higher concentrations paralyses the sense of smell, so that the warning smell of rotten eggs ceases to protect. Carry out the whole stripping stage in an efficient fume cupboard.; The nitrogen outlet from the scrubber must end in the fume cupboard; over the 30 min of stripping the stream of 40 l/h carries the whole stripped amount of hydrogen sulfide.; The colour reagent contains 200 ml of concentrated sulfuric acid per litre — prepare it by carefully adding the acid to the suspension in water and cool it before making up to the mark.; N,N-dimethyl-1,4-phenylenediammonium chloride is sensitising and irritant — weigh it wearing gloves, in a fume cupboard, avoiding dust.; Sodium sulfide is corrosive and in contact with acids releases hydrogen sulfide — do not acidify the stock solutions outside the apparatus and do not pour them into the drain.; Pressure filtration is carried out under nitrogen up to 2 bar — use vessels intended for work under pressure and a shield; nitrogen in a closed room displaces oxygen.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number ISO 10530.