⚗️ Anionic surfactants in water and wastewater — the methylene blue index (MBAS): extraction with chloroform and measurement of the absorbance at 650 nm
In short
In a separating funnel 100 ml of the sample (20–200 µg of MBAS) is made alkaline with a pH 10 buffer (10 ml), 5 ml of neutral methylene blue solution and 15 ml of chloroform are added and it is shaken for at least 1 min; the chloroform layer is washed in a second funnel with 110 ml of water and 5 ml of acidic methylene blue, filtered into a 50 ml flask, the extraction is repeated and the absorbance is measured at 650 nm against chloroform. The test is performed according to PN-EN 903:2002 (EN 903:1993; ISO 7875-1:1984 MOD); Range: 0.1–5.0 mg/l MBAS (as sodium dodecylbenzenesulfonate); limit of detection about 0.05 mg/l; 20–200 µg of MBAS in the portion. The procedure comprises 7 steps; it is used for: Municipal and industrial wastewater (laundries, cosmetics, detergents) — the reference method for anionic surfactants under Dz.U. 2019 poz. 1311, Effluents from treatment plants — assessment of the primary biodegradation of surfactants on a laboratory and a technical scale, Surface waters (rivers, lakes) — monitoring of pollution by detergents; with sublation where the matrix is unknown.
At a glance
- Standard: PN-EN 903:2002 (EN 903:1993; ISO 7875-1:1984 MOD)
- Category: Physicochemistry
- Procedure steps: 7
- Range: 0.1–5.0 mg/l MBAS (as sodium dodecylbenzenesulfonate); limit of detection about 0.05 mg/l; 20–200 µg of MBAS in the portion
- Measurement: absorbance of the chloroform phase at 650 nm, 10 mm cell, against chloroform
- Extraction: 100 ml of sample + 10 ml of pH 10 buffer + 5 ml of neutral blue + 15 ml of chloroform; shaking for at least 1 min; washing with 110 ml of water + 5 ml of acidic blue; the extraction is repeated (2 × 10 ml of chloroform)
Overview
PN-EN 903:2002 (EN 903:1993, based on a modified ISO 7875-1:1984) specifies a spectrometric method for determining anionic surfactants by measuring the methylene blue index (MBAS) in the aquatic environment. Natural and synthetic anionic surface-active substances are determined as substances active towards methylene blue — a sum parameter called the MBAS index. The method (its scope described after the newer edition, ISO 7875-1:1996) applies to drinking water, surface water and wastewater, for example for assessing the primary biodegradation of surfactants in test systems with natural or synthetic wastewater, on a laboratory and on a technical scale; in the effluents of municipal treatment plants the MBAS index covers not only synthetic but, to a considerable extent, also natural anionic surface-active substances. The concentration range is 0.1–5.0 mg/l and the limit of detection is about 0.05 mg/l for standard surfactants in distilled water. The 2019 regulation (Dz.U. 2019 poz. 1311 [Journal of Laws 2019, item 1311]) names PN-EN 903 (molecular absorption spectrophotometry) as the reference method for anionic surfactants in wastewater.
The determination consists in the formation of ion pairs between the methylene blue cation and the surfactant anion, which can be extracted quantitatively with chloroform; what is determined above all are organic sulfates and sulfonates, but positive interferences may occur (organic sulfates, sulfonates, carboxylates, phenols, inorganic anions such as cyanate, nitrate, thiocyanate and sulfide) as well as negative ones (cationic substances — quaternary ammonium compounds, proteins — that form complexes with the anionic surfactants); it is therefore more correct to describe the result as the MBAS index than as "anionic surfactants". Samples are not taken through a layer of foam; the glass bottles are cleaned with methanol; short storage at 4 °C, longer storage with a preservative (1 % of 40 % formaldehyde for up to 4 days, saturation with chloroform for up to 8 days); in surface waters of unknown composition the surfactants can be separated from other active substances by gas stripping (sublation) in an apparatus with ethyl acetate.
Method principle
The formation of the methylene blue salts of the anionic surfactants in an alkaline medium, the extraction of these salts with chloroform and the acid treatment of the chloroform solution; the removal of interferences by extracting the surfactant–methylene blue complex from the alkaline solution and shaking it with an acidic solution of the blue. The absorbance of the separated organic phase is measured at the absorption maximum of 650 nm and evaluated against a calibration curve. The standard is the methyl ester of dodecylbenzenesulfonic acid (the tetrapropylene type, relative molecular mass 340; 400–450 mg is saponified for 1 h under a reflux condenser in 50 ml of ethanolic NaOH, neutralised with H₂SO₄ 0.5 mol/l against phenolphthalein and diluted to 1000 ml; stable for at least 6 months; the sodium salts of dodecane-1-sulfonic acid, dodecane-1-sulfuric acid or dioctylsulfosuccinic acid are also allowed), and the MBAS index is calculated as sodium dodecylbenzenesulfonate. Reagents: neutral methylene blue solution (0.350 g/l, prepared at least 24 h before use, stable for at least 2 weeks; the absorbance of the blank chloroform phase must be ≤ 0.2 per 10 mm, otherwise it is purified by extraction: 200 ml of buffer and 200 ml of chloroform per 100 ml of the solution, shaking for 30 s, running off the chloroform layer and rinsing the aqueous layer with 60 ml of chloroform, twice), acidic methylene blue solution (0.350 g in 500 ml of water + 6.50 ml of H₂SO₄ of 1.84 g/ml, made up to 1000 ml; blank ≤ 0.02 per 10 mm), pH 10 buffer (24 g of NaHCO₃ + 27 g of anhydrous Na₂CO₃ per 1000 ml; for very hard waters, alternatively an alkaline borate solution of equal volumes of 0.05 mol/l Na₂B₄O₇·10H₂O and 0.1 mol/l NaOH), chloroform, ethanol 95 %, methanol, H₂SO₄ 0.5 mol/l, ethanolic NaOH and phenolphthalein. Apparatus: a spectrometer able to measure at 650 nm with 10 mm cells, separating funnels (e.g. 250 ml), volumetric flasks of 50 ml, glass-fibre or soft cellulose filters and — optionally — a gas stripping (sublation) apparatus.
Applications
- Municipal and industrial wastewater (laundries, cosmetics, detergents) — the reference method for anionic surfactants under Dz.U. 2019 poz. 1311
- Effluents from treatment plants — assessment of the primary biodegradation of surfactants on a laboratory and a technical scale
- Surface waters (rivers, lakes) — monitoring of pollution by detergents; with sublation where the matrix is unknown
- Drinking water — confirmation that there are no residues of anionic surfactants (down to 0.05 mg/l)
- Waterworks and environmental laboratories — the classical extraction with spectrophotometric measurement
Key parameters
| Parameter | Value |
|---|---|
| Range | 0.1–5.0 mg/l MBAS (as sodium dodecylbenzenesulfonate); limit of detection about 0.05 mg/l; 20–200 µg of MBAS in the portion |
| Measurement | absorbance of the chloroform phase at 650 nm, 10 mm cell, against chloroform |
| Extraction | 100 ml of sample + 10 ml of pH 10 buffer + 5 ml of neutral blue + 15 ml of chloroform; shaking for at least 1 min; washing with 110 ml of water + 5 ml of acidic blue; the extraction is repeated (2 × 10 ml of chloroform) |
| Methylene blue reagents | neutral 0.350 g/l (blank ≤ 0.2 A per 10 mm); acidic 0.350 g/l + 6.50 ml of H₂SO₄ (blank ≤ 0.02 A per 10 mm); prepared at least 24 h before use |
| Standard | the methyl ester of dodecylbenzenesulfonic acid saponified to the sodium salt; curve from 0 to 8 ml of the 20 µg/ml standard solution per 100 ml |
| Sample | do not take it through the foam; glassware cleaned with methanol; 4 °C for a short time; formaldehyde 1 % (4 days) or saturation with chloroform (8 days) |
Standard
- Standard number
- PN-EN 903:2002 (EN 903:1993; ISO 7875-1:1984 MOD)
- Title (PL)
- Jakość wody — Oznaczanie surfaktantów anionowych przez pomiar indeksu błękitu metylenowego MBAS
- Title (EN)
- Water quality — Determination of anionic surfactants by measurement of the methylene blue index MBAS
Step-by-step procedure
-
Sampling and storage
Take the sample, not through the layer of foam, into a clean glass bottle cleaned with methanol; cool it to 4 °C; for storage longer than 24 h add 1 % (V/V) of 40 % formaldehyde (up to 4 days) or saturate it with chloroform (up to 8 days). Suspended matter can be centrifuged off (surfactant adsorbed on the suspended matter will not be determined).
-
Extraction
Transfer into a separating funnel a measured volume of the sample containing 20–200 µg of MBAS (make up less than 100 ml with water to 100 ml), add 5.0 ml of the neutral methylene blue solution, 10 ml of the pH 10 buffer and 15 ml of chloroform; shake evenly (preferably in the vertical plane) for at least 1 min and leave the phases to separate.
-
Washing the extract
Transfer the chloroform layer as completely as possible into a second funnel containing 110 ml of water and 5 ml of the acidic methylene blue solution; shake for at least 1 min and leave it to separate.
-
Filtration and repetition
Filter the chloroform layer into a 50 ml volumetric flask through a glass-fibre filter moistened with chloroform; repeat the extraction from the first and the second funnel (e.g. twice with 10 ml of chloroform), filtering into the same flask; make up to the mark with chloroform and mix.
-
Measurement
Measure the absorbance of the extract at 650 nm in a 10 mm cell against pure chloroform; for each batch of samples run a blank with 100 ml of distilled water taken through the whole procedure.
-
Calibration
Add 0; 1; 2; 4; 6; 8 ml of the diluted standard solution (e.g. 20 µg/ml of sodium dodecylbenzenesulfonate) to the funnels, make up with water to 100 ml and proceed as with a sample; plot the absorbance-against-concentration curve (including the zero solution).
-
Calculation
Read the MBAS index (as sodium dodecylbenzenesulfonate) from the curve or calculate it: c = (A₁ − A₀)·f₁/V, where A₁ is the absorbance of the sample, A₀ that of the blank, f₁ the calibration factor (the mass of MBAS giving an absorbance of 1.000) and V the volume of the sample; report it in mg/l (µg/l).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Spectrometer (spectrophotometer) able to measure at 650 nm with 10 mm cells | cells cleaned with chloroform | — |
| Separating funnels of 250 ml (2 per determination) | extraction and washing of the extract | — |
| Volumetric flasks of 50 ml, cylinders of 250 ml, pipettes of 5, 10 and 25 ml | dispensing the reagents and collecting the extract | — |
| Glass-fibre filters (or soft cellulose filters) moistened with chloroform, funnel | filtration of the chloroform phase | — |
| Round-bottomed flask with a reflux condenser | saponification of the standard (1 h) | — |
| Gas stripping (sublation) apparatus with ethyl acetate — optional | separation of the surfactants in waters of unknown matrix, work in a fume hood | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Chloroform | — | extraction solvent (the purest available) |
| Methylene blue — neutral solution of 0.350 g/l | — | formation of the ion pair with the surfactant anion |
| Methylene blue — acidic solution (0.350 g/l + 6.50 ml of H₂SO₄/l) | — | washing of the extract, removal of interferences |
| pH 10 buffer (24 g of NaHCO₃ + 27 g of Na₂CO₃/l) or the alkaline borate buffer | — | making the sample alkaline |
| The methyl ester of dodecylbenzenesulfonic acid, ethanolic NaOH, H₂SO₄ 0.5 mol/l, phenolphthalein, ethanol 95 % | — | MBAS standard solution |
| Formaldehyde 40 % or chloroform | — | preservation of the samples (4 or 8 days) |
Health and safety (OHS)
- Chloroform is toxic and suspected of being carcinogenic — carry out all the extractions in a fume hood, wear gloves resistant to solvents, dispose of the chloroform waste
- Ethyl acetate (sublation) — flammable; keep the apparatus in a well-ventilated fume hood
- Formaldehyde (preservation) — work in a fume hood, gloves
- Sulfuric acid and ethanolic NaOH — goggles and gloves
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-EN 903:2002 (EN 903:1993; ISO 7875-1:1984 MOD) — “Water quality — Determination of anionic surfactants by measurement of the methylene blue index MBAS”.
How does this method work?
The formation of the methylene blue salts of the anionic surfactants in an alkaline medium, the extraction of these salts with chloroform and the acid treatment of the chloroform solution; the removal of interferences by extracting the surfactant–methylene blue complex from the alkaline solution and shaking it with an acidic solution of the blue. The absorbance of the separated organic phase is measured at the absorption maximum of 650 nm and evaluated against a calibration curve.
What is the measuring range and accuracy?
Range: 0.1–5.0 mg/l MBAS (as sodium dodecylbenzenesulfonate); limit of detection about 0.05 mg/l; 20–200 µg of MBAS in the portion; Measurement: absorbance of the chloroform phase at 650 nm, 10 mm cell, against chloroform; Extraction: 100 ml of sample + 10 ml of pH 10 buffer + 5 ml of neutral blue + 15 ml of chloroform; shaking for at least 1 min; washing with 110 ml of water + 5 ml of acidic blue; the extraction is repeated (2 × 10 ml of chloroform); Methylene blue reagents: neutral 0.350 g/l (blank ≤ 0.2 A per 10 mm); acidic 0.350 g/l + 6.50 ml of H₂SO₄ (blank ≤ 0.02 A per 10 mm); prepared at least 24 h before use.
How long does the test take?
The procedure comprises 7 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Spectrometer (spectrophotometer) able to measure at 650 nm with 10 mm cells, Separating funnels of 250 ml (2 per determination), Volumetric flasks of 50 ml, cylinders of 250 ml, pipettes of 5, 10 and 25 ml, Glass-fibre filters (or soft cellulose filters) moistened with chloroform, funnel, Round-bottomed flask with a reflux condenser, Gas stripping (sublation) apparatus with ethyl acetate — optional.
Where is this test used?
Municipal and industrial wastewater (laundries, cosmetics, detergents) — the reference method for anionic surfactants under Dz.U. 2019 poz. 1311; Effluents from treatment plants — assessment of the primary biodegradation of surfactants on a laboratory and a technical scale; Surface waters (rivers, lakes) — monitoring of pollution by detergents; with sublation where the matrix is unknown; Drinking water — confirmation that there are no residues of anionic surfactants (down to 0.05 mg/l); Waterworks and environmental laboratories — the classical extraction with spectrophotometric measurement.
What safety precautions apply?
Chloroform is toxic and suspected of being carcinogenic — carry out all the extractions in a fume hood, wear gloves resistant to solvents, dispose of the chloroform waste; Ethyl acetate (sublation) — flammable; keep the apparatus in a well-ventilated fume hood; Formaldehyde (preservation) — work in a fume hood, gloves; Sulfuric acid and ethanolic NaOH — goggles and gloves.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-EN 903.