🌿 Iron oxides in workplace air — flame AAS
In short
Sampling of the respirable (or of the inhalable) fraction of the aerosol of iron oxides on a cellulose ester membrane filter, hot mineralisation of the filter with concentrated nitric acid, preparation of a solution in dilute HNO₃ with a lanthanum buffer and determination of iron by atomic absorption spectrometry in an air–acetylene flame at 248. The test is performed according to PN-Z-04469:2025-02; Measuring range: 0.35–10.4 mg/m³ with V1 = 50 ml, k = 10 — 2013 procedure (0.07–2.1 of the NDS then in force, 5 mg/m³); 2021 procedure: 0.35–10.42 mg/m³ (inhalable fraction) and 0.17–5.21 mg/m³ (respirable fraction). The procedure comprises 7 steps; it is used for: Assessment of occupational exposure to welding fumes and to dusts containing iron oxides — welding shops, grinding shops, metallurgy, ore mines, foundries, Measurements required by the regulation of the Minister of Health on tests and measurements of harmful agents — comparison with the NDS for the inhalable fraction (5 mg/m³) and for the respirable fraction (2.5 mg/m³), Personal sampling in the breathing zone with separation into the fractions of the aerosol.
At a glance
- Standard: PN-Z-04469:2025-02
- Category: Environment
- Procedure steps: 7
- Wavelength / flame: 248.3 nm; air–acetylene (lean); hollow cathode lamp for Fe
- Calibration curve: 0.50; 1.50; 5.00; 10.0; 15.0 µg/ml Fe with the addition of about 0.5 % La; correlation coefficient 1.0000
- Limit of quantification: 0.10 µg/ml in the solution; limit of determination 0.35 mg/m³ for the respirable fraction (720 l) — 2013 procedure; card of the 2025-02 edition: 0.35 mg/m³ for the inhalable fraction and 0.17 mg/m³ for the respirable fraction
Overview
The edition PN-Z-04469:2025-02 replaced PN-Z-04469:2015-10.
Iron oxides — haematite (Fe₂O₃), magnetite (Fe₃O₄), iron(II) oxide — occur in the air of ore mines, in metallurgical processes and above all in welding. Their inhalation has an irritant action; with continuous exposure they are deposited in the lungs, causing iron pneumoconiosis (siderosis) — the oldest known benign pneumoconiosis — and high concentrations (more than 10 mg Fe/m³) may damage the respiratory tract ("welder's lung"). In the regulation on the maximum admissible concentrations (NDS) and intensities (NDN) iron oxides — calculated as Fe — have admissible values laid down for the inhalable fraction (5 mg/m³) and for the respirable fraction (2.5 mg/m³).
The method employing flame atomic absorption spectrometry was worked out at CIOP-PIB as a revision of the standard PN-79/Z-04066.02, which met neither the requirements of PN-EN 482 (a limit of determination at the level of 1/10 of the NDS) nor the principles of personal sampling of PN-Z-04008-7; that procedure became the basis of the standard PN-Z-04469:2015-10 (the next edition 2025-02). In the 2013 procedure, from which the course described below is taken, the limit of determination for the respirable fraction is 0.35 mg/m³ with an air volume of 720 l, and the measuring range of 0.35–10.4 mg/m³ corresponded to 0.07–2.1 of the NDS value in force at that time (5 mg/m³ for the respirable fraction); where there are other iron compounds in the air the sum of the iron is determined. The more recent CIOP-PIB procedure (2021) determines manganese, nickel and iron in both fractions of the aerosol in a single analytical run — for iron in the range of 0.35–10.42 mg/m³ in the inhalable fraction and 0.17–5.21 mg/m³ in the respirable fraction (720 l of air, 10 ml of solution after mineralisation), i.e. about 0.07–2.08 of the NDS values of 2018. The PKN catalogue card of the PN-Z-04469:2025-02 edition (read on 23.09.2026) gives the same lowest determinable concentrations: 0.35 mg/m³ for the inhalable fraction and 0.17 mg/m³ for the respirable fraction. How the course according to the 2025-02 edition differs from the one described below — WE HAVE NOT ESTABLISHED: we do not have the text of the standard.
Method principle
Course according to the CIOP-PIB procedure of 2013 (the basis of the 2015-10 edition). A known volume of air (up to 720 l) is drawn through a cellulose ester membrane filter with a pore diameter of 0.85 µm placed in the head of a cyclone or in another separator of the respirable fraction (for the inhalable fraction — in a suitable sampler). The filter is mineralised in a beaker with 2 ml of concentrated HNO₃ (65 %) on a hot plate at a temperature of about 140 °C almost to dryness (the operation is repeated twice), 4 ml of 1 mol/l HNO₃ is added, after 30 min it is transferred quantitatively with 0.1 mol/l HNO₃ into a 50 ml flask, and 1 ml of that solution is diluted in a 10 ml flask with the addition of 0.5 ml of lanthanum chloride solution (about 10 % La) as a spectral buffer. Iron is determined in a (lean) air–acetylene flame at a wavelength of 248.3 nm, against a calibration curve of 0.50–15.0 µg/ml prepared on the day of the analysis, the blank from a clean filter being subtracted. The concentration in the air: X = (c − c0)·V1·k/V [mg/m³], where V1 = 50 ml, k = 10, V — the volume of air in litres. The presence of Mn, Ni, Cr and Cu (50–200 µg/ml) does not affect the result.
Applications
- Assessment of occupational exposure to welding fumes and to dusts containing iron oxides — welding shops, grinding shops, metallurgy, ore mines, foundries
- Measurements required by the regulation of the Minister of Health on tests and measurements of harmful agents — comparison with the NDS for the inhalable fraction (5 mg/m³) and for the respirable fraction (2.5 mg/m³)
- Personal sampling in the breathing zone with separation into the fractions of the aerosol
- Determination of the sum of the iron in the aerosol where other iron compounds occur together with the oxides
- Parallel determination of Mn, Ni and Fe from one sample in the CIOP-PIB procedure (2021)
Key parameters
| Parameter | Value |
|---|---|
| Wavelength / flame | 248.3 nm; air–acetylene (lean); hollow cathode lamp for Fe |
| Calibration curve | 0.50; 1.50; 5.00; 10.0; 15.0 µg/ml Fe with the addition of about 0.5 % La; correlation coefficient 1.0000 |
| Limit of quantification | 0.10 µg/ml in the solution; limit of determination 0.35 mg/m³ for the respirable fraction (720 l) — 2013 procedure; card of the 2025-02 edition: 0.35 mg/m³ for the inhalable fraction and 0.17 mg/m³ for the respirable fraction |
| Measuring range | 0.35–10.4 mg/m³ with V1 = 50 ml, k = 10 — 2013 procedure (0.07–2.1 of the NDS then in force, 5 mg/m³); 2021 procedure: 0.35–10.42 mg/m³ (inhalable fraction) and 0.17–5.21 mg/m³ (respirable fraction) |
| Precision / total uncertainty | 5.3 % / 11.0 % (2013 procedure: 5.29 % / 11.02 %); recovery from the filter 0.99 (required 1.00 ± 0.05) |
| NDS (list as worded by Dz.U. 2026 poz. 447 [Journal of Laws 2026, item 447], item 530 — values as in Dz.U. 2018 poz. 1286, item 504) | inhalable fraction 5 mg/m³ (NDSCh 10 mg/m³); respirable fraction 2.5 mg/m³ (NDSCh 5 mg/m³); calculated as Fe |
Standard
- Standard number
- PN-Z-04469:2025-02
- Title (PL)
- Ochrona czystości powietrza — Oznaczanie tlenków żelaza na stanowiskach pracy metodą płomieniowej absorpcyjnej spektrometrii atomowej
- Title (EN)
- Air purity protection — Determination of iron oxides at workplaces by flame atomic absorption spectrometry
Step-by-step procedure
-
Sampling
Calibrate the system for the sampling of the respirable fraction with a flow calibrator for the flow rate required by the manufacturer of the cyclone. In the breathing zone of the worker (the principles of PN-Z-04008-7) draw up to 720 l of air through the filter placed in the head of the cyclone. Where the inhalable fraction is determined take a parallel sample in a sampler of that fraction.
-
Mineralisation
Place the filter in a 50 ml beaker, add 2 ml of concentrated HNO₃ and heat it on the hot plate (about 140 °C) almost to dryness; repeat twice with 2 ml of the acid. After cooling add 4 ml of 1 mol/l HNO₃ and leave it for about 30 min. Transfer it quantitatively with 5–6 portions of 0.1 mol/l HNO₃ into a 50 ml flask and make up to the mark. In parallel mineralise a clean filter (the blank).
-
Solution for the test
Into a 10 ml flask transfer 1 ml of the solution after the mineralisation, add 0.5 ml of lanthanum chloride solution and make up with 0.1 mol/l HNO₃. At concentrations outside the range change the dilution factor, retaining the addition of lanthanum.
-
Calibration curve
On the day of the analysis prepare the working solutions of 0.50; 1.50; 5.00; 10.0 and 15.0 µg/ml Fe (0.5 ml of lanthanum chloride per 10 ml) and a zero sample. Make three measurements of the absorbance of each solution at 248.3 nm in a lean air–acetylene flame, zeroing the spectrometer with 0.1 mol/l HNO₃; prepare the curve or use the automatic calibration of the instrument.
-
Measurement and calculation
Measure the absorbance of the solution for the test and of the blank under the same conditions, read the concentrations c and c0 from the curve. Calculate X = (c − c0)·V1·k/V in mg/m³ (V1 = 50 ml, k = 10, V — the volume of air in l).
-
Check of the effectiveness of the mineralisation
Onto six filters apply 600 µl each of the 10 mg/ml Fe solution, dry them until the following day, mineralise them as the samples and compare them with comparison solutions prepared without mineralisation; the factor wm = (Cd − Co)/Cp should be 1.00 ± 0.05 — otherwise change the manner of the mineralisation.
-
Assessment of the exposure
Determine the index of exposure for the working shift and compare it with the NDS appropriate to the fraction (5 mg/m³ inhalable, 2.5 mg/m³ respirable).
Required equipment and apparatus
| Equipment | Example | Indicative price |
|---|---|---|
| Atomic absorption spectrometer with an air–acetylene burner | with a hollow cathode lamp for iron (in the validation a Thermo SOLAAR M); the parameters of the lamp, of the slit and of the height of the burner chosen to suit the instrument | — |
| Suction pump (personal sampler) | a constant volume flow rate as recommended by the manufacturer of the cyclone; a flow calibrator for the check before the measurement | — |
| Cyclone or separator of the respirable fraction with a head for the filter | or a sampler of the inhalable fraction where that fraction is determined | — |
| Cellulose ester membrane filters | pore diameter 0.85 µm (e.g. Pragopor 4) | — |
| Hot plate | mineralisation of the filters in 50 ml beakers at a temperature of about 140 °C, under a fume cupboard | — |
| Borosilicate glass and polyethylene vessels, automatic pipettes | flasks of 50 ml and 10 ml; the vessels washed successively with detergent, warm water, 1 mol/l HNO₃, distilled water and deionised water | — |
Reagents, media and consumables
| Reagent | CAS | Details |
|---|---|---|
| Concentrated nitric acid 65 % (ρ = 1.39 g/ml), analytical grade | — | mineralisation of the filters |
| Nitric acid, solutions of 1 mol/l and 0.1 mol/l | — | dissolution of the residue after the mineralisation, making up of the flasks, zeroing of the spectrometer |
| Lanthanum chloride solution (26.8 g of LaCl₃·7H₂O in 73.2 ml of 0.1 mol/l HNO₃; about 10 % La) | — | spectral buffer in the standard solutions and in the test solutions |
| Basic standard solution of iron for AAS 1 mg/ml (e.g. GUM) | — | intermediate solution 50 µg/ml (stable for 14 days) and working solutions 0.50–15.0 µg/ml |
| Standard solution of iron 10 mg/ml | — | testing of the effectiveness of the mineralisation (of the recovery) |
| Dissolved acetylene of class A (PN-C-84905) | — | combustible gas for the flame |
| Doubly distilled or deionised water (18.2 MΩ·cm) | — | preparation of all the solutions |
Health and safety (OHS)
- Carry out all the operations with concentrated nitric acid in protective clothing and goggles and under an efficiently working fume cupboard — the hot mineralisation evolves nitrogen oxides
- Acetylene: the gas installation and the burner in accordance with the instructions for the spectrometer; check of leak-tightness, protection against flash-back of the flame
- Collect the residues of the solutions of the reagents and of the standards in marked containers and hand them over to authorised firms for disposal
Frequently asked questions
Which standard describes this test?
The test is performed according to PN-Z-04469:2025-02 — “Air purity protection — Determination of iron oxides at workplaces by flame atomic absorption spectrometry”.
How does this method work?
Course according to the CIOP-PIB procedure of 2013 (the basis of the 2015-10 edition). A known volume of air (up to 720 l) is drawn through a cellulose ester membrane filter with a pore diameter of 0.
What is the measuring range and accuracy?
Wavelength / flame: 248.3 nm; air–acetylene (lean); hollow cathode lamp for Fe; Calibration curve: 0.50; 1.50; 5.00; 10.0; 15.0 µg/ml Fe with the addition of about 0.5 % La; correlation coefficient 1.0000; Limit of quantification: 0.10 µg/ml in the solution; limit of determination 0.35 mg/m³ for the respirable fraction (720 l) — 2013 procedure; card of the 2025-02 edition: 0.35 mg/m³ for the inhalable fraction and 0.17 mg/m³ for the respirable fraction; Measuring range: 0.35–10.4 mg/m³ with V1 = 50 ml, k = 10 — 2013 procedure (0.07–2.1 of the NDS then in force, 5 mg/m³); 2021 procedure: 0.35–10.42 mg/m³ (inhalable fraction) and 0.17–5.21 mg/m³ (respirable fraction).
How long does the test take?
The procedure comprises 7 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.
What equipment is required?
Atomic absorption spectrometer with an air–acetylene burner, Suction pump (personal sampler), Cyclone or separator of the respirable fraction with a head for the filter, Cellulose ester membrane filters, Hot plate, Borosilicate glass and polyethylene vessels, automatic pipettes.
Where is this test used?
Assessment of occupational exposure to welding fumes and to dusts containing iron oxides — welding shops, grinding shops, metallurgy, ore mines, foundries; Measurements required by the regulation of the Minister of Health on tests and measurements of harmful agents — comparison with the NDS for the inhalable fraction (5 mg/m³) and for the respirable fraction (2.5 mg/m³); Personal sampling in the breathing zone with separation into the fractions of the aerosol; Determination of the sum of the iron in the aerosol where other iron compounds occur together with the oxides; Parallel determination of Mn, Ni and Fe from one sample in the CIOP-PIB procedure (2021).
What safety precautions apply?
Carry out all the operations with concentrated nitric acid in protective clothing and goggles and under an efficiently working fume cupboard — the hot mineralisation evolves nitrogen oxides; Acetylene: the gas installation and the burner in accordance with the instructions for the spectrometer; check of leak-tightness, protection against flash-back of the flame; Collect the residues of the solutions of the reagents and of the standards in marked containers and hand them over to authorised firms for disposal.
Which laboratory can perform this test?
The test is performed by laboratories accredited to ISO/IEC 17025. On LabCoda you can find them by the standard number PN-Z-04469.