📏 Quality assurance of AMS — QAL2 calibration, QAL3 control and the annual AST test

Metrology PN-EN 14181

In short

The quality assurance procedures for automated measuring systems for emissions: QAL2 — determination of the calibration function and of the variability of the AMS from parallel measurements by the standard reference method (SRM); QAL3 — ongoing control of the drift and the precision at zero and at span with control charts; AST — an annual functional test and a limited number of parallel measurements. The test is performed according to PN-EN 14181:2015-02 (EN 14181:2014); Certification range: ≤ 1.5 × the daily ELV (incineration plants), ≤ 2.5 × (large combustion plants). The procedure comprises 6 steps; it is used for: Calibration and validation (QAL2) of continuous emission monitoring systems for SO₂, NOx, CO, dust, O₂, H₂O, HCl, NH₃, Hg and others after installation, Annual control (AST) and five-yearly full calibration of continuous measurement systems required by the regulation on measurements of emissions, Ongoing control by the operator (QAL3) — control charts of the drift and the precision at zero and at span.

At a glance

  • Standard: PN-EN 14181:2015-02 (EN 14181:2014)
  • Category: Metrology
  • Procedure steps: 6
  • Quality levels: QAL1 (suitability, EN 15267-3 / EN ISO 14956), QAL2 (calibration and variability), QAL3 (drift and precision), AST (annual test)
  • Calibration function: linear y = a + b·x from SRM vs AMS (dust: a quadratic one permissible in accordance with EN 13284-2)
  • Variability: standard deviation sD of the differences Di between the SRM and the calibrated AMS; test kv (χ²) against σ₀

Overview

EN 14181 describes the quality assurance procedures needed to demonstrate that an automated measuring system (AMS) installed to measure emissions to air meets the requirements for the uncertainty of the measured values arising from the law (EU directives, national provisions) or from the competent authorities. Three levels of quality assurance are defined: QAL1 — the suitability of the AMS for the task (in accordance with EN 15267-3 and EN ISO 14956, with a calculation of the total uncertainty from the performance characteristics), QAL2 — validation after installation, QAL3 — control during operation, and the annual surveillance test (AST). The standard specifies: the QAL2 procedure for calibrating the AMS and determining the variability of the measured values, in order to demonstrate the suitability of the AMS after installation; the QAL3 procedure for maintaining and demonstrating the quality of the results in normal operation by checking the agreement of the zero and span characteristics with those established in QAL1; the AST procedure assessing whether the AMS is working correctly and whether the calibration function and the variability remain valid. It is applied after the certification of the AMS in accordance with the EN 15267 series; the QAL2 and AST procedures are carried out by testing laboratories, QAL3 — by the operator of the plant. The 2014 edition replaced EN 14181:2004. In Poland the regulation on the measurement of the quantity of emissions requires that continuous measurement systems be subject to procedures in accordance with PN-EN 14181, including at least once a year a control by parallel measurements with the reference methods (dust, SO₂, NOx, CO, O₂, water vapour), the full calibration and validation procedure for new systems, for existing ones at least once every five years and after major changes, and that the confidence intervals of a single result be determined in accordance with that standard.

Method principle

The AMS must have had its suitability for the task (the parameter and the composition of the gas) demonstrated in QAL1 — new plants require certification in accordance with EN 15267-1, -2, -3; EN 15267-3 requires the total uncertainty from the test to be at least 25 % below the maximum permissible one, in order to leave a margin for the influence of the installation. QAL2: on a correctly installed and commissioned AMS a series of parallel measurements is carried out with the standard reference method (SRM) in the same sampling plane; from the pairs of results a linear calibration function is determined (for dust a quadratic one is permissible in accordance with EN 13284-2) as well as the variability — the standard deviation of the differences between the SRM and the calibrated AMS — and it is compared by a test with the maximum permissible uncertainty (σ₀ derived from the provisions, e.g. σ₀ = P·E/1.96, where the uncertainty is given as half of the 95 % confidence interval as a percentage P of the permissible value E). QAL2 is repeated periodically, after a major change in the operation of the plant, after a failure of the AMS or where the law requires it. QAL3: periodic checks of the zero and the span with reference materials (as in the repeatability tests of QAL1) and assessment of the results with CUSUM or EWMA control charts against the uncertainty components from QAL1; where necessary, adjustment or servicing. AST: functional tests and a limited number of parallel SRM measurements checking whether the uncertainty and the calibration function from the previous QAL2 still hold. The standard does not cover the uncertainty of the data acquisition and handling system (DAHS); it distinguishes the certification range (≤ 1.5 × the daily ELV for incineration plants, ≤ 2.5 × for large combustion plants), the calibration range and the measuring range.

Applications

Key parameters

ParameterValue
Quality levelsQAL1 (suitability, EN 15267-3 / EN ISO 14956), QAL2 (calibration and variability), QAL3 (drift and precision), AST (annual test)
Calibration functionlinear y = a + b·x from SRM vs AMS (dust: a quadratic one permissible in accordance with EN 13284-2)
Variabilitystandard deviation sD of the differences Di between the SRM and the calibrated AMS; test kv (χ²) against σ₀
Maximum uncertaintyfrom the provisions; σ₀ = P·E/1.96 for an uncertainty given as half of the 95 % confidence interval (P % of the ELV)
Certification range≤ 1.5 × the daily ELV (incineration plants), ≤ 2.5 × (large combustion plants)
QAL3 control chartsCUSUM or EWMA — drift and change of precision vs the uncertainty components from QAL1
Frequency (PL, regulation of 2021)parallel measurements at least once a year; full calibration and validation: new systems, existing ones ≥ once every 5 years, after major changes

Standard

Standard number
PN-EN 14181:2015-02 (EN 14181:2014)
Title (PL)
Emisja ze źródeł stacjonarnych — Zapewnienie jakości automatycznych systemów pomiarowych
Title (EN)
Stationary source emissions — Quality assurance of automated measuring systems

Step-by-step procedure

  1. QAL1 — suitability

    Make sure that the AMS is certified in accordance with EN 15267-1/-2/-3 for the parameter and the composition of the gas and that its total uncertainty from QAL1 meets the requirements of the provisions with a margin of 25 %; for existing non-certified systems apply annex H.2.

  2. QAL2 — parallel measurements

    After installation and commissioning have a laboratory carry out a series of parallel SRM measurements (in the same sampling plane, under conditions covering the operating range of the plant) with simultaneous recording of the AMS signal, after a functional check of the AMS.

  3. QAL2 — calibration function and variability

    From the pairs (xi, yi) determine the calibration function (the best estimates of a and b), calculate the differences Di and their standard deviation sD, carry out the variability test against σ₀; establish the calibration range (ys,min – ys,max). A non-compliant result → corrective actions and repetition.

  4. QAL3 — ongoing control

    The operator periodically (at intervals no longer than the maintenance interval) checks the zero and the span with reference materials, plots the results on CUSUM or EWMA charts and compares them with the limits from QAL1; on an exceedance — adjustment or maintenance.

  5. AST — annual test

    Each year the laboratory carries out functional tests of the AMS and a limited number of parallel SRM measurements; it checks whether the variability is within the criteria and whether the calibration function from QAL2 remains valid; in the event of non-compliance — a new QAL2.

  6. Repetition of QAL2 and documentation

    Repeat QAL2 periodically (in Poland at least once every 5 years), after a major change in the operation of the plant, after a failure or a repair of the AMS; document the calibration function, the variability, the confidence intervals of a single result and all the tests.

Required equipment and apparatus

EquipmentExampleIndicative price
Certified AMS (EN 15267) with a sampling and conditioning systemextractive or in-situ; peripheral AMS (H₂O, T, p, O₂) for the conversions to standard conditions—
Sets of reference methods (SRM) for the parallel measurementsdust EN 13284-1, SO₂ EN 14791, NOx EN 14792, O₂ EN 14789, CO EN 15058, H₂O EN 14790, HCl EN 1911, TVOC EN 12619, NH₃ EN ISO 21877, Hg EN 13211—
Reference materials for zero and spantest gases or simulating filters/devices of known characteristics (QAL3, AST)—
Software for regression, variability tests and CUSUM/EWMA chartsdetermination of a, b, sD, kv—
Data acquisition and handling system (DAHS)outside the scope of the assessment of the standard, but it affects the quality of the results—

Health and safety (OHS)

Frequently asked questions

Which standard describes this test?

The test is performed according to PN-EN 14181:2015-02 (EN 14181:2014) — “Stationary source emissions — Quality assurance of automated measuring systems”.

How does this method work?

The AMS must have had its suitability for the task (the parameter and the composition of the gas) demonstrated in QAL1 — new plants require certification in accordance with EN 15267-1, -2, -3; EN 15267-3 requires the total uncertainty from the test to be at least 25 % below the maximum permissible one, in order to leave a margin for the influence of the installation. QAL2: on a correctly installed and commissioned AMS a series of parallel measurements is carried out with the standard reference method (SRM) in the same sampling plane; from the pairs of results a linear calibration function is determined (for dust a quadratic one is permissible in accordance with EN 13284-2) as well as the variability — the standard deviation of the differences between the SRM and the calibrated AMS — and it is compared by a test with the maximum permissible uncertainty (σ₀ derived from the provisions, e.

What is the measuring range and accuracy?

Quality levels: QAL1 (suitability, EN 15267-3 / EN ISO 14956), QAL2 (calibration and variability), QAL3 (drift and precision), AST (annual test); Calibration function: linear y = a + b·x from SRM vs AMS (dust: a quadratic one permissible in accordance with EN 13284-2); Variability: standard deviation sD of the differences Di between the SRM and the calibrated AMS; test kv (χ²) against σ₀; Maximum uncertainty: from the provisions; σ₀ = P·E/1.96 for an uncertainty given as half of the 95 % confidence interval (P % of the ELV).

How long does the test take?

The procedure comprises 6 steps. The standard does not give a duration for every stage — the laboratory's own procedure decides.

What equipment is required?

Certified AMS (EN 15267) with a sampling and conditioning system, Sets of reference methods (SRM) for the parallel measurements, Reference materials for zero and span, Software for regression, variability tests and CUSUM/EWMA charts, Data acquisition and handling system (DAHS).

Where is this test used?

Calibration and validation (QAL2) of continuous emission monitoring systems for SO₂, NOx, CO, dust, O₂, H₂O, HCl, NH₃, Hg and others after installation; Annual control (AST) and five-yearly full calibration of continuous measurement systems required by the regulation on measurements of emissions; Ongoing control by the operator (QAL3) — control charts of the drift and the precision at zero and at span; Determination of the confidence intervals of a single result of a continuous measurement for comparison with the emission standards; Assessment of existing, non-certified AMS (annex H.2) and of peripheral systems (temperature, pressure, O₂, water vapour).